RadioBanter

RadioBanter (https://www.radiobanter.com/)
-   Antenna (https://www.radiobanter.com/antenna/)
-   -   Coaxial folded dipole (was: Natural balun/Antenna on 9/26/2004) (https://www.radiobanter.com/antenna/2410-coaxial-folded-dipole-natural-balun-antenna-9-26-2004-a.html)

John Smith October 4th 04 09:19 PM

Coaxial folded dipole (was: Natural balun/Antenna on 9/26/2004)
 
Okay, then, I will present data measured this day for this antenna:

http://www.sophisticatedsolutions.us...d%20Dipole.jpg

This is shown in "Antennas for All Applications" on page 820, figure 23-17
(a).

I built the antenna wholly from RG58. The center conductor of the right half
is not connected at either end. It is 14.375 inches wide and averages a
little less than .5 inches between the centers of the top and bottom
conductors. Where the coax is shown exiting the antenna, is a female,
flangeless, chassis mount, BNC connector so that I can replace the antenna
with a short.

My test set up is a VHF oscillator, a vector voltmeter, and a Narda dual
directional coupler. I use a 66 inch piece of RG58 from the output of the
directional coupler to go to the antenna. The short circuits I use are the
best I could make from BNC connectors. The 50 Ohm load I used for
comparative measurements is one of those used for network terminators. Yes,
I am aware they are not instrumentation quality, but it's what I have.

For a given frequency, I replace the antenna with the short and adjust the
amplitude of the oscillator and the controls of the vector voltmeter so that
the reference channel (A) is 10 mV and the phase is 180 degrees. I record
channel B's amplitude. I then remove the short and connect the antenna. I
then read and record channels A, channel B, and the phase. From these data I
calculate the impedance (per HP's AN 77-3, thanks to Wes Stewart).

The first item measured is the 50 Ohm terminator. I also measured it at the
conclusion of the tests to see if there were any differences and there were
none.

Here are the results computed from the data:

Freq (MHz) Impedance (Ohms)

410 46.4 + 6.0i (50 Ohm terminator)

380 9.7 - 12.5
390 3.5 - 5.7
400 5.1 + 1.3i
410 5.1 + 6.5i
415 4.0 + 10.0i
425 2.5 + 15.7i

This is surprising to me and doesn't make a whole lot of sense. For one
thing, I would have expected the impedance to vary wildly over the range
shown. For another, the low impedance seems, well, really low.

Is the trend of the data as shown to be expected? Well, maybe the reactive
part? The real part seems to make a little sense except at the extremes.

Can I trust this data to be even approximately right? I mean, can I now say
that the antenna in question is maybe nice due to the natural balun but I
might as well forget it as a simple antenna because to the low impedance?

Or, should I say this is utter nonsense, the antenna is probably okay, it's
just my equipment, setup, or lack of knowledge giving erroneous data?

Your opinions are welcome.

John (KD5YI)



Wes Stewart October 4th 04 10:34 PM

On Mon, 4 Oct 2004 15:19:41 -0500, "John Smith"
wrote:

|Okay, then, I will present data measured this day for this antenna:
|
|http://www.sophisticatedsolutions.us...d%20Dipole.jpg
|
|This is shown in "Antennas for All Applications" on page 820, figure 23-17
|(a).
|
|I built the antenna wholly from RG58. The center conductor of the right half
|is not connected at either end. It is 14.375 inches wide and averages a
|little less than .5 inches between the centers of the top and bottom
|conductors. Where the coax is shown exiting the antenna, is a female,
|flangeless, chassis mount, BNC connector so that I can replace the antenna
|with a short.
|
|My test set up is a VHF oscillator, a vector voltmeter, and a Narda dual
|directional coupler. I use a 66 inch piece of RG58 from the output of the
|directional coupler to go to the antenna. The short circuits I use are the
|best I could make from BNC connectors. The 50 Ohm load I used for
|comparative measurements is one of those used for network terminators. Yes,
|I am aware they are not instrumentation quality, but it's what I have.
|
|For a given frequency, I replace the antenna with the short and adjust the
|amplitude of the oscillator and the controls of the vector voltmeter so that
|the reference channel (A) is 10 mV and the phase is 180 degrees. I record
|channel B's amplitude. I then remove the short and connect the antenna. I
|then read and record channels A, channel B, and the phase. From these data I
|calculate the impedance (per HP's AN 77-3, thanks to Wes Stewart).
|
|The first item measured is the 50 Ohm terminator. I also measured it at the
|conclusion of the tests to see if there were any differences and there were
|none.
|
|Here are the results computed from the data:
|
|Freq (MHz) Impedance (Ohms)
|
|410 46.4 + 6.0i (50 Ohm terminator)
|
|380 9.7 - 12.5
|390 3.5 - 5.7
|400 5.1 + 1.3i
|410 5.1 + 6.5i
|415 4.0 + 10.0i
|425 2.5 + 15.7i
|
|This is surprising to me and doesn't make a whole lot of sense. For one
|thing, I would have expected the impedance to vary wildly over the range
|shown. For another, the low impedance seems, well, really low.
|
|Is the trend of the data as shown to be expected? Well, maybe the reactive
|part? The real part seems to make a little sense except at the extremes.
|
|Can I trust this data to be even approximately right? I mean, can I now say
|that the antenna in question is maybe nice due to the natural balun but I
|might as well forget it as a simple antenna because to the low impedance?
|
|Or, should I say this is utter nonsense, the antenna is probably okay, it's
|just my equipment, setup, or lack of knowledge giving erroneous data?
|
|Your opinions are welcome.

John,

First of all you are to be commended for running these experiments.
Without a bit more study of your situation I can't comment too much
but I wanted to throw out a couple of quick ideas.

1) Do you have the VVM probes terminated in 50 ohm?

2) If you don't have the line stretcher as in Fig 7 of the note, are
you recalibrating at each test frequency?

3) How well is your signal source terminated, in other words do you
know its source match?

Each of these things can affect the outcome.

More later,


Wes

John Smith October 4th 04 10:45 PM

Hi, Wes -


"Wes Stewart" wrote in message
...
On Mon, 4 Oct 2004 15:19:41 -0500, "John Smith"
wrote:


John,

First of all you are to be commended for running these experiments.
Without a bit more study of your situation I can't comment too much
but I wanted to throw out a couple of quick ideas.

1) Do you have the VVM probes terminated in 50 ohm?



Yes. They came witht the kit and are called terminators. They are 50 Ohms
each.


2) If you don't have the line stretcher as in Fig 7 of the note, are
you recalibrating at each test frequency?



I do not have the stretcher. I recalibrate at every frequency change.


3) How well is your signal source terminated, in other words do you
know its source match?



I only know that the signal source is an HP 3200B. It directly feeds the
Narda dual directional coupler through a few feet of RG58.


Each of these things can affect the outcome.

More later,


Wes


Thanks, Wes.

John



Richard Clark October 4th 04 11:34 PM

On Mon, 4 Oct 2004 15:19:41 -0500, "John Smith"
wrote:

Here are the results computed from the data:

Freq (MHz) Impedance (Ohms)

410 46.4 + 6.0i (50 Ohm terminator)

380 9.7 - 12.5
390 3.5 - 5.7
400 5.1 + 1.3i
410 5.1 + 6.5i
415 4.0 + 10.0i
425 2.5 + 15.7i


Hi John,

How about the raw data? How about a detailed description of the NARDA
coupler? It should have a calibration plate affixed to the side of it
with freq vs. coupling marks (generally pretty close).

I would note you have a considerable SWR, but this was expected going
into the test (however, in an inverse proportion, which may be
meaningful here).

73's
Richard Clark, KB7QHC

Wes Stewart October 4th 04 11:50 PM

On Mon, 4 Oct 2004 16:45:31 -0500, "John Smith"
wrote:

|Hi, Wes -
|
|
|"Wes Stewart" wrote in message
.. .
| On Mon, 4 Oct 2004 15:19:41 -0500, "John Smith"
| wrote:
|
| John,
|
| First of all you are to be commended for running these experiments.
| Without a bit more study of your situation I can't comment too much
| but I wanted to throw out a couple of quick ideas.
|
| 1) Do you have the VVM probes terminated in 50 ohm?
|
|
|Yes. They came witht the kit and are called terminators. They are 50 Ohms
|each.

Okay.
|
|
| 2) If you don't have the line stretcher as in Fig 7 of the note, are
| you recalibrating at each test frequency?
|
|
|I do not have the stretcher. I recalibrate at every frequency change.
|
Understood.
|
| 3) How well is your signal source terminated, in other words do you
| know its source match?
|
|
|I only know that the signal source is an HP 3200B. It directly feeds the
|Narda dual directional coupler through a few feet of RG58.

If I remember that correctly the '3200 is nothing but a p-p oscillator
and a waveguide-below-cutoff probe. If your VVM reference probe
readings are changing much between frequencies and/or
calibration/measurement, try a 6 or 10 dB pad right on the generator
output and see what happens.

When you're calibrating using a short, the source Z has really got to
be nailed down.

Wes


Tam/WB2TT October 5th 04 12:50 AM


"John Smith" wrote in message
...
Okay, then, I will present data measured this day for this antenna:

http://www.sophisticatedsolutions.us...d%20Dipole.jpg

This is shown in "Antennas for All Applications" on page 820, figure 23-17
(a).

I built the antenna wholly from RG58. The center conductor of the right
half is not connected at either end. It is 14.375 inches wide and averages
a little less than .5 inches between the centers of the top and bottom
conductors. Where the coax is shown exiting the antenna, is a female,
flangeless, chassis mount, BNC connector so that I can replace the antenna
with a short.

My test set up is a VHF oscillator, a vector voltmeter, and a Narda dual
directional coupler. I use a 66 inch piece of RG58 from the output of the
directional coupler to go to the antenna. The short circuits I use are the
best I could make from BNC connectors. The 50 Ohm load I used for
comparative measurements is one of those used for network terminators.
Yes, I am aware they are not instrumentation quality, but it's what I
have.

For a given frequency, I replace the antenna with the short and adjust the
amplitude of the oscillator and the controls of the vector voltmeter so
that the reference channel (A) is 10 mV and the phase is 180 degrees. I
record channel B's amplitude. I then remove the short and connect the
antenna. I then read and record channels A, channel B, and the phase. From
these data I calculate the impedance (per HP's AN 77-3, thanks to Wes
Stewart).

The first item measured is the 50 Ohm terminator. I also measured it at
the conclusion of the tests to see if there were any differences and there
were none.

Here are the results computed from the data:

Freq (MHz) Impedance (Ohms)

410 46.4 + 6.0i (50 Ohm terminator)

380 9.7 - 12.5
390 3.5 - 5.7
400 5.1 + 1.3i
410 5.1 + 6.5i
415 4.0 + 10.0i
425 2.5 + 15.7i
........................................

John,
Discounting the absolute values, the numbers seem to make sense, except for
the 9.7. Might it have been 2.7? There seems to be resonance at around 400.
The impedance goes more reactive in both directions from there, and the real
part goes down monotonically, except for the 9.7

I looked at the picture, and it is not 100% obvious to me what gets
connected at the balanced point. Just for kicks, I am going to try that,
somewhere within the range of an MFJ269.

Tam/WB2TT



John Smith October 5th 04 01:18 AM


"Wes Stewart" wrote in message
...
On Mon, 4 Oct 2004 16:45:31 -0500, "John Smith"


If I remember that correctly the '3200 is nothing but a p-p oscillator
and a waveguide-below-cutoff probe. If your VVM reference probe
readings are changing much between frequencies and/or
calibration/measurement, try a 6 or 10 dB pad right on the generator
output and see what happens.



The 3200B is the oscillator they use in the HP AN 77-3 you sent to me. I'm
using it the same way they did except for the stretcher and a Narda (rather
than HP) coupler. However, I'll try to run a test and determine how much
difference there is using a pad.


When you're calibrating using a short, the source Z has really got to
be nailed down.

Wes



Thanks.

John



John Smith October 5th 04 01:30 AM


"Richard Clark" wrote in message
...
On Mon, 4 Oct 2004 15:19:41 -0500, "John Smith"
wrote:

Here are the results computed from the data:

Freq (MHz) Impedance (Ohms)

410 46.4 + 6.0i (50 Ohm terminator)

380 9.7 - 12.5
390 3.5 - 5.7
400 5.1 + 1.3i
410 5.1 + 6.5i
415 4.0 + 10.0i
425 2.5 + 15.7i


Hi John,

How about the raw data? How about a detailed description of the NARDA
coupler? It should have a calibration plate affixed to the side of it
with freq vs. coupling marks (generally pretty close).


The Narda coupler has no calibration plate. It says model 30611, serial no.
4235. It also appears to have been made for Motorola, as their part number
appears on it. One end says "BTS" and the opposite end says "ANT". At the
BTS end, on the side, there is a connector which says -30 dB. Similarly on
the antenna end.


Raw data:

Reference Measurement
Freq A1 B1 ?1 A2 B2 ?2
415 1 0.76 180 0.82 0.535 158
410 1 0.77 180 0.905 0.57 165
400 1 0.79 180 1.08 0.695 177
390 1 0.81 180 1.02 0.72 -167
425 1 0.743 180 1.06 0.72 145
380 1 0.815 180 0.86 0.485 -151
410 1 0.749 180 0.535 0.029 118
425 1 0.695 180 1.04 0.695 143


I would note you have a considerable SWR, but this was expected going
into the test (however, in an inverse proportion, which may be
meaningful here).

73's
Richard Clark, KB7QHC




John Smith October 5th 04 01:42 AM


"Tam/WB2TT" wrote in message
...

"John Smith" wrote in message
...
Okay, then, I will present data measured this day for this antenna:

http://www.sophisticatedsolutions.us...d%20Dipole.jpg

This is shown in "Antennas for All Applications" on page 820, figure
23-17 (a).

I built the antenna wholly from RG58. The center conductor of the right
half is not connected at either end. It is 14.375 inches wide and
averages a little less than .5 inches between the centers of the top and
bottom conductors. Where the coax is shown exiting the antenna, is a
female, flangeless, chassis mount, BNC connector so that I can replace
the antenna with a short.

My test set up is a VHF oscillator, a vector voltmeter, and a Narda dual
directional coupler. I use a 66 inch piece of RG58 from the output of the
directional coupler to go to the antenna. The short circuits I use are
the best I could make from BNC connectors. The 50 Ohm load I used for
comparative measurements is one of those used for network terminators.
Yes, I am aware they are not instrumentation quality, but it's what I
have.

For a given frequency, I replace the antenna with the short and adjust
the amplitude of the oscillator and the controls of the vector voltmeter
so that the reference channel (A) is 10 mV and the phase is 180 degrees.
I record channel B's amplitude. I then remove the short and connect the
antenna. I then read and record channels A, channel B, and the phase.
From these data I calculate the impedance (per HP's AN 77-3, thanks to
Wes Stewart).

The first item measured is the 50 Ohm terminator. I also measured it at
the conclusion of the tests to see if there were any differences and
there were none.

Here are the results computed from the data:

Freq (MHz) Impedance (Ohms)

410 46.4 + 6.0i (50 Ohm terminator)

380 9.7 - 12.5
390 3.5 - 5.7
400 5.1 + 1.3i
410 5.1 + 6.5i
415 4.0 + 10.0i
425 2.5 + 15.7i
........................................

John,
Discounting the absolute values, the numbers seem to make sense, except
for the 9.7. Might it have been 2.7? There seems to be resonance at around
400. The impedance goes more reactive in both directions from there, and
the real part goes down monotonically, except for the 9.7

I looked at the picture, and it is not 100% obvious to me what gets
connected at the balanced point. Just for kicks, I am going to try that,
somewhere within the range of an MFJ269.

Tam/WB2TT



Hi, Tam -

I will try to repeat the test at that frequency.

By the balanced point, I assume you mean at the bottom center. It is a
female BNC connector, facing downward. A halfwave length of RG58 goes off
the left side and folds. The coax is soldered in normal fashion to the
connector. Another halfwave piece of RG58 goes off the right side and folds.
The center conductor of the right side piece is not connected on either end.
The shield of the right side coax is soldered to the shell of the BNC and
the two pices of coax is joined as shown in the figure.

If by the balanced point you meant at the top center of the figure, the
center conductor only of the left side coax is soldered to the shield only
of the right side coax.

If this description is not clear, let me know and I'll try again. I would
take a picture and make it available, but I'm afraid it would only confuse
due to lack of detail.

Thanks,
John



John Smith October 5th 04 02:08 AM


"Tam/WB2TT" wrote in message
...

"John Smith" wrote in message
...
Okay, then, I will present data measured this day for this antenna:

http://www.sophisticatedsolutions.us...d%20Dipole.jpg

This is shown in "Antennas for All Applications" on page 820, figure
23-17 (a).

I built the antenna wholly from RG58. The center conductor of the right
half is not connected at either end. It is 14.375 inches wide and
averages a little less than .5 inches between the centers of the top and
bottom conductors. Where the coax is shown exiting the antenna, is a
female, flangeless, chassis mount, BNC connector so that I can replace
the antenna with a short.

My test set up is a VHF oscillator, a vector voltmeter, and a Narda dual
directional coupler. I use a 66 inch piece of RG58 from the output of the
directional coupler to go to the antenna. The short circuits I use are
the best I could make from BNC connectors. The 50 Ohm load I used for
comparative measurements is one of those used for network terminators.
Yes, I am aware they are not instrumentation quality, but it's what I
have.

For a given frequency, I replace the antenna with the short and adjust
the amplitude of the oscillator and the controls of the vector voltmeter
so that the reference channel (A) is 10 mV and the phase is 180 degrees.
I record channel B's amplitude. I then remove the short and connect the
antenna. I then read and record channels A, channel B, and the phase.
From these data I calculate the impedance (per HP's AN 77-3, thanks to
Wes Stewart).

The first item measured is the 50 Ohm terminator. I also measured it at
the conclusion of the tests to see if there were any differences and
there were none.

Here are the results computed from the data:

Freq (MHz) Impedance (Ohms)

410 46.4 + 6.0i (50 Ohm terminator)

380 9.7 - 12.5
390 3.5 - 5.7
400 5.1 + 1.3i
410 5.1 + 6.5i
415 4.0 + 10.0i
425 2.5 + 15.7i
........................................

John,
Discounting the absolute values, the numbers seem to make sense, except
for the 9.7. Might it have been 2.7? There seems to be resonance at around
400. The impedance goes more reactive in both directions from there, and
the real part goes down monotonically, except for the 9.7


Well, measured again, at 380 MHz I get 0.9 - 4.4i.

I looked at the picture, and it is not 100% obvious to me what gets
connected at the balanced point. Just for kicks, I am going to try that,
somewhere within the range of an MFJ269.

Tam/WB2TT



This is getting discouraging.

John




John Smith October 5th 04 02:58 AM


"Wes Stewart" wrote in message
...
On Mon, 4 Oct 2004 16:45:31 -0500, "John Smith"
wrote:


Freq (MHz) Impedance (Ohms)

410 46.4 + 6.0i (50 Ohm terminator)

380 9.7 - 12.5
390 3.5 - 5.7
400 5.1 + 1.3i
410 5.1 + 6.5i
415 4.0 + 10.0i
425 2.5 + 15.7i


| 3) How well is your signal source terminated, in other words do you
| know its source match?
|
|
|I only know that the signal source is an HP 3200B. It directly feeds the
|Narda dual directional coupler through a few feet of RG58.

If I remember that correctly the '3200 is nothing but a p-p oscillator
and a waveguide-below-cutoff probe. If your VVM reference probe
readings are changing much between frequencies and/or
calibration/measurement, try a 6 or 10 dB pad right on the generator
output and see what happens.

When you're calibrating using a short, the source Z has really got to
be nailed down.

Wes



Okay. I repeated the test using an HP 355C attenuator set for 10 dB and at
400 MHz got 4 + 3i on the antenna. I also checked my 50 Ohm network
terminator with this setup and it measured 44 + 4i. The data are different,
but they're not an order of magnitude different, at least.

So, although my measurements aren't repeatable, they are sloppily
consistent. That is, although I can't say exactly what the antenna impedance
is with confidence, I am beginning to believe that it really is very low in
impedance. Am I drawing an erroneous conclusion too early? I can wait a
little longer to draw an erroneous conclusion.

John




Tam/WB2TT October 5th 04 03:05 AM


"John Smith" wrote in message
...

"Tam/WB2TT" wrote in message
...

"John Smith" wrote in message
...
Okay, then, I will present data measured this day for this antenna:

http://www.sophisticatedsolutions.us...d%20Dipole.jpg

This is shown in "Antennas for All Applications" on page 820, figure
23-17 (a).

I built the antenna wholly from RG58. The center conductor of the right
half is not connected at either end. It is 14.375 inches wide and
averages a little less than .5 inches between the centers of the top and
bottom conductors. Where the coax is shown exiting the antenna, is a
female, flangeless, chassis mount, BNC connector so that I can replace
the antenna with a short.

My test set up is a VHF oscillator, a vector voltmeter, and a Narda dual
directional coupler. I use a 66 inch piece of RG58 from the output of
the directional coupler to go to the antenna. The short circuits I use
are the best I could make from BNC connectors. The 50 Ohm load I used
for comparative measurements is one of those used for network
terminators. Yes, I am aware they are not instrumentation quality, but
it's what I have.

For a given frequency, I replace the antenna with the short and adjust
the amplitude of the oscillator and the controls of the vector voltmeter
so that the reference channel (A) is 10 mV and the phase is 180 degrees.
I record channel B's amplitude. I then remove the short and connect the
antenna. I then read and record channels A, channel B, and the phase.
From these data I calculate the impedance (per HP's AN 77-3, thanks to
Wes Stewart).

The first item measured is the 50 Ohm terminator. I also measured it at
the conclusion of the tests to see if there were any differences and
there were none.

Here are the results computed from the data:

Freq (MHz) Impedance (Ohms)

410 46.4 + 6.0i (50 Ohm terminator)

380 9.7 - 12.5
390 3.5 - 5.7
400 5.1 + 1.3i
410 5.1 + 6.5i
415 4.0 + 10.0i
425 2.5 + 15.7i
........................................

John,
Discounting the absolute values, the numbers seem to make sense, except
for the 9.7. Might it have been 2.7? There seems to be resonance at
around 400. The impedance goes more reactive in both directions from
there, and the real part goes down monotonically, except for the 9.7

I looked at the picture, and it is not 100% obvious to me what gets
connected at the balanced point. Just for kicks, I am going to try that,
somewhere within the range of an MFJ269.

Tam/WB2TT



Hi, Tam -

I will try to repeat the test at that frequency.

By the balanced point, I assume you mean at the bottom center. It is a
female BNC connector, facing downward. A halfwave length of RG58 goes off
the left side and folds. The coax is soldered in normal fashion to the
connector. Another halfwave piece of RG58 goes off the right side and
folds. The center conductor of the right side piece is not connected on
either end. The shield of the right side coax is soldered to the shell of
the BNC and the two pices of coax is joined as shown in the figure.

If by the balanced point you meant at the top center of the figure, the
center conductor only of the left side coax is soldered to the shield only
of the right side coax.

If this description is not clear, let me know and I'll try again. I would
take a picture and make it available, but I'm afraid it would only confuse
due to lack of detail.

Thanks,
John

John, This turned out simpler than I thought. I did a quick and dirty test
with a folded dipole made up of two 3 foot pieces of RG58, and a feedline
1/2 WL at 160 MHz. Unfortunately, this puts me too close to the upper limit
of the MFJ, but you can see what happens. BTW, I meant the top for the
balanced point. Turns out it makes no difference whether the center
conductor of the left side is connected to the center conductor, or the
shield, of the right side.


My numbers:

F R X
145 30 89
150 10 60
155 ? 7? 14? This looks like 165
160 8 25
165 7 13
170.25 6 0
175 5 14

I would say that if anything, my numbers make less sense than yours. I also
got a reading of 1+j0 at about 65 MHz. Don't know what that means, other
than a dead short. I will hook it up as a regular folded dipole, and see if
I get anything like 300 Ohms. Probably tomorrow.

Was this claimed to be a 50 Ohm antenna?

Tam/WB2TT



Richard Clark October 5th 04 08:04 AM

On Mon, 4 Oct 2004 19:30:19 -0500, "John Smith"
wrote:

The Narda coupler has no calibration plate. It says model 30611, serial no.
4235. It also appears to have been made for Motorola, as their part number
appears on it.


Hi John,

This is not a NARDA model number according to their own catalogue,
although I see it described as NARDA in more than one ebay auction. I
would say this is a special run for Motorola (which is probably their
contract number with NARDA).

However, all ebay auctions list this as an 960 MHz device.

Typically, directional couplers are within their nominal ratings only
over a octave range and some of those octaves from within their
catalogue are 500 MHz to 1 GHz. Others are 450 MHz to 900 MHz. Some
are listed as 0.05 GHz to 1 GHz, but the coupling is VERY MUCH
different than nominal outside the octave range (by as much as 10 - 15
dB). Generally, you don't suffer this much variation near the octave
band edges, but they do track off from their otherwise flat response.

It looks like you need to measure the coupling directly (at both
ports) across your frequencies of interest. The coupling factor is
not so important as is the ports tracking. I will discuss the raw and
finished data separately.

73's
Richard Clark, KB7QHC

Richard Clark October 5th 04 08:35 AM

On Mon, 4 Oct 2004 19:30:19 -0500, "John Smith"
wrote:
400 5.1 + 1.3i

Reference Measurement
Freq A1 B1 ?1 A2 B2 ?2
400 1 0.79 180 1.08 0.695 177


Hi John,

Well, from the two results above, and referencing my copy of Appl.
Note 77-3, Page 7, section "Measuring Rho 100 to 1000 MHz," there are
a number of issues here.

Your B1/A1 is quite off the mark (but certainly correctable, afterall,
that is the purpose of its measurement).

|Rho| = B2·A1/A2·B1 = 0.695·1/1.08·0.79 = 0.815

Casting the magnitude and angle onto a Smith Chart would suggest your
5.1 + 1.3i Ohms is close enough given your data.

The only kicker is port tracking, but I have a hunch that probably is
not an issue.

This bears further consideration. Too bad Kraus did not choose to
elaborate.

73's
Richard Clark, KB7QHC

[email protected] October 5th 04 01:20 PM

On Mon, 4 Oct 2004 15:19:41 -0500, "John Smith"
wrote:

Okay, then, I will present data measured this day for this antenna:

http://www.sophisticatedsolutions.us...d%20Dipole.jpg

This is shown in "Antennas for All Applications" on page 820, figure 23-17
(a).


Just an off the cuff remark and a possible clue.

It's not clear from the picture or explanation if the center
conductor is continious from the right half of the antenna to the
left. I understand the center conductor does not connect to the
right half at the T.

For the case that it does continue:
Looking that the picture, if the right side were parallel to the left
I'd call that a 1/4 wave 1:1 balun with a shorted load connected
rather than an antenna. That would result in a significant reactance
or a short depending on frequency. The sorted load would be the
continued centor conductor in the right half.

For the case that it does Not continue:
I'd call that a 1/4 wave 1:1 balun with an open load connected rather
than an antenna. That would result in a significant reactance again.

There is a third case:
A connect dot is missing at the junction of the center conductor where
it meets the shield of the right loop (top center). If the lengths
were ~1/4 wave for each side then the impedence at the center would
be high for the left center conductor and the right shield and that
would likely be a tuneable folded dipole.

Did I miss something about the antenna design?

As drawn it looks like an attempt to take a parallel line balun (coax
with a 1/4wave 1:1 balun) and make it serve as a radiator. There is
detail missing one possible dimensions and other connections.

FYI: if you used one of those cheap eithernet Tees to create the
junction, I've found them to be very poor at UHF. Use a good quality
one with TFE insulation and test it seperately first.

Allison

John Smith October 5th 04 02:40 PM


wrote in message
...
On Mon, 4 Oct 2004 15:19:41 -0500, "John Smith"
wrote:

Okay, then, I will present data measured this day for this antenna:

http://www.sophisticatedsolutions.us...d%20Dipole.jpg

This is shown in "Antennas for All Applications" on page 820, figure 23-17
(a).


Just an off the cuff remark and a possible clue.

It's not clear from the picture or explanation if the center
conductor is continious from the right half of the antenna to the
left. I understand the center conductor does not connect to the
right half at the T.


The center conductor in the right half is not used at all. The right half
should actually be a piece of coax with the center conductor removed.

For the case that it does continue:
Looking that the picture, if the right side were parallel to the left
I'd call that a 1/4 wave 1:1 balun with a shorted load connected
rather than an antenna. That would result in a significant reactance
or a short depending on frequency. The sorted load would be the
continued centor conductor in the right half.

For the case that it does Not continue:
I'd call that a 1/4 wave 1:1 balun with an open load connected rather
than an antenna. That would result in a significant reactance again.

There is a third case:
A connect dot is missing at the junction of the center conductor where
it meets the shield of the right loop (top center). If the lengths
were ~1/4 wave for each side then the impedence at the center would
be high for the left center conductor and the right shield and that
would likely be a tuneable folded dipole.

Did I miss something about the antenna design?



The center conductor of the left side is connected to the shield of the
right side. The shield of the left side is not connected at the top center.


As drawn it looks like an attempt to take a parallel line balun (coax
with a 1/4wave 1:1 balun) and make it serve as a radiator. There is
detail missing one possible dimensions and other connections.

FYI: if you used one of those cheap eithernet Tees to create the
junction, I've found them to be very poor at UHF. Use a good quality
one with TFE insulation and test it seperately first.

Allison


I wound up not using a T. I simply soldered all connections to a BNC
connector at the bottom center of the figure.

John



John Smith October 5th 04 03:00 PM


"Tam/WB2TT" wrote in message
...

"John Smith" wrote in message
...

"Tam/WB2TT" wrote in message
...

"John Smith" wrote in message
...
Okay, then, I will present data measured this day for this antenna:

http://www.sophisticatedsolutions.us...d%20Dipole.jpg

This is shown in "Antennas for All Applications" on page 820, figure
23-17 (a).

I built the antenna wholly from RG58. The center conductor of the right
half is not connected at either end. It is 14.375 inches wide and
averages a little less than .5 inches between the centers of the top
and bottom conductors. Where the coax is shown exiting the antenna, is
a female, flangeless, chassis mount, BNC connector so that I can
replace the antenna with a short.

My test set up is a VHF oscillator, a vector voltmeter, and a Narda
dual directional coupler. I use a 66 inch piece of RG58 from the output
of the directional coupler to go to the antenna. The short circuits I
use are the best I could make from BNC connectors. The 50 Ohm load I
used for comparative measurements is one of those used for network
terminators. Yes, I am aware they are not instrumentation quality, but
it's what I have.

For a given frequency, I replace the antenna with the short and adjust
the amplitude of the oscillator and the controls of the vector
voltmeter so that the reference channel (A) is 10 mV and the phase is
180 degrees. I record channel B's amplitude. I then remove the short
and connect the antenna. I then read and record channels A, channel B,
and the phase. From these data I calculate the impedance (per HP's AN
77-3, thanks to Wes Stewart).

The first item measured is the 50 Ohm terminator. I also measured it at
the conclusion of the tests to see if there were any differences and
there were none.

Here are the results computed from the data:

Freq (MHz) Impedance (Ohms)

410 46.4 + 6.0i (50 Ohm terminator)

380 9.7 - 12.5
390 3.5 - 5.7
400 5.1 + 1.3i
410 5.1 + 6.5i
415 4.0 + 10.0i
425 2.5 + 15.7i
........................................
John,
Discounting the absolute values, the numbers seem to make sense, except
for the 9.7. Might it have been 2.7? There seems to be resonance at
around 400. The impedance goes more reactive in both directions from
there, and the real part goes down monotonically, except for the 9.7

I looked at the picture, and it is not 100% obvious to me what gets
connected at the balanced point. Just for kicks, I am going to try that,
somewhere within the range of an MFJ269.

Tam/WB2TT



Hi, Tam -

I will try to repeat the test at that frequency.

By the balanced point, I assume you mean at the bottom center. It is a
female BNC connector, facing downward. A halfwave length of RG58 goes off
the left side and folds. The coax is soldered in normal fashion to the
connector. Another halfwave piece of RG58 goes off the right side and
folds. The center conductor of the right side piece is not connected on
either end. The shield of the right side coax is soldered to the shell of
the BNC and the two pices of coax is joined as shown in the figure.

If by the balanced point you meant at the top center of the figure, the
center conductor only of the left side coax is soldered to the shield
only of the right side coax.

If this description is not clear, let me know and I'll try again. I would
take a picture and make it available, but I'm afraid it would only
confuse due to lack of detail.

Thanks,
John

John, This turned out simpler than I thought. I did a quick and dirty test
with a folded dipole made up of two 3 foot pieces of RG58, and a feedline
1/2 WL at 160 MHz. Unfortunately, this puts me too close to the upper
limit of the MFJ, but you can see what happens. BTW, I meant the top for
the balanced point. Turns out it makes no difference whether the center
conductor of the left side is connected to the center conductor, or the
shield, of the right side.


My numbers:

F R X
145 30 89
150 10 60
155 ? 7? 14? This looks like 165
160 8 25
165 7 13
170.25 6 0
175 5 14

I would say that if anything, my numbers make less sense than yours. I
also got a reading of 1+j0 at about 65 MHz. Don't know what that means,
other than a dead short. I will hook it up as a regular folded dipole, and
see if I get anything like 300 Ohms. Probably tomorrow.

Was this claimed to be a 50 Ohm antenna?

Tam/WB2TT



Hey, Tam!

This is great information! According to the dimensions, it should be
resonant at 173 MHz. (Close enough... I subtracted 1 inch on each side for
the loop). Your 6 Ohms is close enough to my 5 Ohms to satisfy me that I'm
not getting useless measurements.

I wouldn't worry about the 65 MHz reading, as that's so far away from
resonance that it is probably acting like a shorted half wave transmission
line or some other kind of crazy network.

No, there was no statement given about the antenna impedance. That's why I
started this thread. I was interested in the antenna because:

1. It is at DC ground.
2. It is a half wave, giving a nice omnidirectional pattern if mounted
vertically.
3. No balun is needed as it is inherent.

However, it appears to be not worthwhile after all, owing to its low
feedpoint impedance.

I really appreciate you taking the time to perform your experiment. I'm now
convinced that the antenna is next to worthless.

Thanks,
John
(KD5YI)



John Smith October 5th 04 03:10 PM


"Richard Clark" wrote in message
...
On Mon, 4 Oct 2004 19:30:19 -0500, "John Smith"
wrote:
400 5.1 + 1.3i

Reference Measurement
Freq A1 B1 ?1 A2 B2 ?2
400 1 0.79 180 1.08 0.695 177


Hi John,

Well, from the two results above, and referencing my copy of Appl.
Note 77-3, Page 7, section "Measuring Rho 100 to 1000 MHz," there are
a number of issues here.

Your B1/A1 is quite off the mark (but certainly correctable, afterall,
that is the purpose of its measurement).

|Rho| = B2·A1/A2·B1 = 0.695·1/1.08·0.79 = 0.815

Casting the magnitude and angle onto a Smith Chart would suggest your
5.1 + 1.3i Ohms is close enough given your data.

The only kicker is port tracking, but I have a hunch that probably is
not an issue.

This bears further consideration. Too bad Kraus did not choose to
elaborate.

73's
Richard Clark, KB7QHC



I don't understand. Please enumerate the issues. What do you mean B1/A1 is
off the mark? If they are off the mark, how can the Z be close enough?

Are you saying that I have calculated the Z correctly from the data and you
think port tracking is not at fault? What further consideration?

You're right. Too bad Kraus didn't tell the feedpoint impedance. I suspect I
would not have embarked on this folly if he had. The antenna may have its
uses elsewhere, but I don't need the headaches of matching 5 Ohms at 440
MHz.

John



Richard Clark October 5th 04 05:25 PM

On Tue, 5 Oct 2004 09:10:19 -0500, "John Smith"
wrote:

I don't understand. Please enumerate the issues. What do you mean B1/A1 is
off the mark?


Hi John,

You should get 1.0 @ 180° (the definition of a short).

If they are off the mark, how can the Z be close enough?


Because what you DID measure, was used as a correction factor per:
|Rho| = B2·A1/A2·B1 = 0.695·1/1.08·0.79 = 0.815
(or you skipped that step)

Are you saying that I have calculated the Z correctly from the data and you
think port tracking is not at fault?


Well, that is really your job to confirm or deny. There is very
little I can accomplish short of that.

What further consideration?


I suppose I could visit my Engineering Library. I will be on campus
for my Nanotechnology seminar today anyway.

73's
Richard Clark, KB7QHC

John Smith October 5th 04 06:26 PM


"Richard Clark" wrote in message
...
On Tue, 5 Oct 2004 09:10:19 -0500, "John Smith"
wrote:

I don't understand. Please enumerate the issues. What do you mean B1/A1 is
off the mark?


Hi John,

You should get 1.0 @ 180° (the definition of a short).


Well, that's not possible when feeding a length of RG58 at 400 MHz, is it?
Remember, I said that there was about a 5 foot length of RG58 between the
directional coupler and the load. How can one get 1.0 reflected to the
coupler when the load is a short? That requires zero loss coax.


If they are off the mark, how can the Z be close enough?


Because what you DID measure, was used as a correction factor per:
|Rho| = B2·A1/A2·B1 = 0.695·1/1.08·0.79 = 0.815
(or you skipped that step)


I did that. As far as I can determine, I did it like the HP application note
said to do it.

Thanks for your comments.

John



Tam/WB2TT October 5th 04 07:07 PM

John,
I had breakfast with some friends this morning. One of them has a PHD in EE,
specializing in antenna design. He thinks the antenna should work, but
suggested changing the spacing between the upper and lower conductors. He
also told me to look for multiple resonances. I just now tried that, with
the spacing essentially 0. What I got was a new resonance at 165 MHz, with
Z=18 + j0. . Interestingly, I now seem to have hit parallel resonance, like
you did, and R goes down either side of 165. In fact, accross 100 -150 MHz
the Z is 0 -jX. When I was messing around with spacing, at one time I got
202 + j0 at 129 MHz; but that is way off calculated frequency. I probably
should not be doing this inside, as it is not entirely repeatable.

I also tried a longer piece of coax (RG8X, Vp=.75) that gave me a 5 foot
folded dipole. That should be resonant at about 93 MHz. I clearly got
multiple resonances:

F R X
92.5 3 j3
100 11 j44
110 80 j181
120 4 j30
127 4 j1
160 53 j14 SWR=1.3
170 11 j0

The feedline was also 5 feet, since I did not know what a wavelengt was
going to be. This looks like it wants to be a 160 MHz antenna, instead of
93. Lastly, I tried the 5 foot antennawith a 2 inch feedline.Resonance was
at 156 MHz. If I get a chance, I will try it outdoors tomorrow.

Tam/WB2TT



Wes Stewart October 5th 04 09:06 PM

On Mon, 4 Oct 2004 20:58:07 -0500, "John Smith"
wrote:

|
|"Wes Stewart" wrote in message
.. .
| On Mon, 4 Oct 2004 16:45:31 -0500, "John Smith"
| wrote:
|
|Freq (MHz) Impedance (Ohms)
|
|410 46.4 + 6.0i (50 Ohm terminator)
|
|380 9.7 - 12.5
|390 3.5 - 5.7
|400 5.1 + 1.3i
|410 5.1 + 6.5i
|415 4.0 + 10.0i
|425 2.5 + 15.7i
|
|
| | 3) How well is your signal source terminated, in other words do you
| | know its source match?
| |
| |
| |I only know that the signal source is an HP 3200B. It directly feeds the
| |Narda dual directional coupler through a few feet of RG58.
|
| If I remember that correctly the '3200 is nothing but a p-p oscillator
| and a waveguide-below-cutoff probe. If your VVM reference probe
| readings are changing much between frequencies and/or
| calibration/measurement, try a 6 or 10 dB pad right on the generator
| output and see what happens.
|
| When you're calibrating using a short, the source Z has really got to
| be nailed down.
|
| Wes
|
|
|Okay. I repeated the test using an HP 355C attenuator set for 10 dB and at
|400 MHz got 4 + 3i on the antenna. I also checked my 50 Ohm network
|terminator with this setup and it measured 44 + 4i. The data are different,
|but they're not an order of magnitude different, at least.
|
|So, although my measurements aren't repeatable, they are sloppily
|consistent. That is, although I can't say exactly what the antenna impedance
|is with confidence, I am beginning to believe that it really is very low in
|impedance. Am I drawing an erroneous conclusion too early? I can wait a
|little longer to draw an erroneous conclusion.

First of all, neglecting the feed method, the antenna is a simple
folded dipole. In free space, or an approximation thereof, it should
have a feedpoint Z of about 300 ohm. (See the ARRL Antenna book for a
description of why this is so under "Special Antenna Types", p.2-32 in
the 17th edition) In the presence of other (non-resonant) objects, it
may differ from this but not a whole lot.

In theory, the "natural balun" doesn't change the impedance of the
feedpoint whatsoever. By "feedpoint" I mean the gap between the ends
of the folded element, not the "tee" connection opposite.

At the outside of the tee connection, the voltage is zero thus this
point can be grounded, connected to the boom in a Yagi, etc. without
upsetting anything. Likewise the coax feeder can be introduced here
and run through one side of the element without upsetting anything
either.

But, a nominal 300 ohm load is terminating a 50 ohm line, so the usual
transforming effects are in play. The input Z of an arbitrary length
line is---well, arbitrary. If the line is many wavelengths long, then
when the frequency is changed, the long lines effect kicks in and the
input Z is going to vary rapidly with respect to frequency.

Second. I believe that you need to determine the parameters of your
directional coupler. As Richard pointed out, your B1/A1 numbers are
pretty unstable.

So here's what I recommend. First verify that the "A" and "B" probes
read the same thing when connected to the same source. Then put your
pad right at the input connector of the coupler. Terminate the
reverse port and connect your VVM "A" probe to the ouput connector and
the "B" probe to the forward port.

The ratio reading is the forward coupling factor of the directional
coupler. Vary the frequency and see how this changes and note some
values. Move the "B" probe to the reverse port and terminate the
forward port. Note the readings at the same frequency.

Reverse the input and output ports and repeat the measurements.
Ideally, the data sets will track closely. If they don't then you
have a problem. Serious differences might indicate damage to the
internal terminations. This assumes that this is a true dual coupler
and not single line coupler with the termination applied to the unused
port externally.

If the numbers are consistant, then you can determine the directivity
by computing the ratio between the two readings on a given port when
the coupler is reversed.

I'm going to stop here and assume you understand the consquences of
poor directivity on measurement accuracy. If you don't then I can
expound further later.

Wes

John Smith October 5th 04 09:23 PM


"Tam/WB2TT" wrote in message
...
John,
I had breakfast with some friends this morning. One of them has a PHD in
EE, specializing in antenna design. He thinks the antenna should work, but
suggested changing the spacing between the upper and lower conductors. He
also told me to look for multiple resonances. I just now tried that, with
the spacing essentially 0. What I got was a new resonance at 165 MHz, with
Z=18 + j0. . Interestingly, I now seem to have hit parallel resonance,
like you did, and R goes down either side of 165. In fact, accross
100 -150 MHz the Z is 0 -jX. When I was messing around with spacing, at
one time I got 202 + j0 at 129 MHz; but that is way off calculated
frequency. I probably should not be doing this inside, as it is not
entirely repeatable.

I also tried a longer piece of coax (RG8X, Vp=.75) that gave me a 5 foot
folded dipole. That should be resonant at about 93 MHz. I clearly got
multiple resonances:

F R X
92.5 3 j3
100 11 j44
110 80 j181
120 4 j30
127 4 j1
160 53 j14 SWR=1.3
170 11 j0

The feedline was also 5 feet, since I did not know what a wavelengt was
going to be. This looks like it wants to be a 160 MHz antenna, instead of
93. Lastly, I tried the 5 foot antennawith a 2 inch feedline.Resonance was
at 156 MHz. If I get a chance, I will try it outdoors tomorrow.

Tam/WB2TT



Thanks for the hard work, Tam. I'm not sure I know what to make of all this,
but it appears that the antenna is not what I thought it would be.

Thanks again.

John



John Smith October 5th 04 09:26 PM


"Wes Stewart" wrote in message
...
On Mon, 4 Oct 2004 20:58:07 -0500, "John Smith"
wrote:

|
|"Wes Stewart" wrote in message
.. .
| On Mon, 4 Oct 2004 16:45:31 -0500, "John Smith"
| wrote:
|
|Freq (MHz) Impedance (Ohms)
|
|410 46.4 + 6.0i (50 Ohm terminator)
|
|380 9.7 - 12.5
|390 3.5 - 5.7
|400 5.1 + 1.3i
|410 5.1 + 6.5i
|415 4.0 + 10.0i
|425 2.5 + 15.7i
|
|
| | 3) How well is your signal source terminated, in other words do you
| | know its source match?
| |
| |
| |I only know that the signal source is an HP 3200B. It directly feeds
the
| |Narda dual directional coupler through a few feet of RG58.
|
| If I remember that correctly the '3200 is nothing but a p-p oscillator
| and a waveguide-below-cutoff probe. If your VVM reference probe
| readings are changing much between frequencies and/or
| calibration/measurement, try a 6 or 10 dB pad right on the generator
| output and see what happens.
|
| When you're calibrating using a short, the source Z has really got to
| be nailed down.
|
| Wes
|
|
|Okay. I repeated the test using an HP 355C attenuator set for 10 dB and
at
|400 MHz got 4 + 3i on the antenna. I also checked my 50 Ohm network
|terminator with this setup and it measured 44 + 4i. The data are
different,
|but they're not an order of magnitude different, at least.
|
|So, although my measurements aren't repeatable, they are sloppily
|consistent. That is, although I can't say exactly what the antenna
impedance
|is with confidence, I am beginning to believe that it really is very low
in
|impedance. Am I drawing an erroneous conclusion too early? I can wait a
|little longer to draw an erroneous conclusion.

First of all, neglecting the feed method, the antenna is a simple
folded dipole. In free space, or an approximation thereof, it should
have a feedpoint Z of about 300 ohm. (See the ARRL Antenna book for a
description of why this is so under "Special Antenna Types", p.2-32 in
the 17th edition) In the presence of other (non-resonant) objects, it
may differ from this but not a whole lot.

In theory, the "natural balun" doesn't change the impedance of the
feedpoint whatsoever. By "feedpoint" I mean the gap between the ends
of the folded element, not the "tee" connection opposite.

At the outside of the tee connection, the voltage is zero thus this
point can be grounded, connected to the boom in a Yagi, etc. without
upsetting anything. Likewise the coax feeder can be introduced here
and run through one side of the element without upsetting anything
either.

But, a nominal 300 ohm load is terminating a 50 ohm line, so the usual
transforming effects are in play. The input Z of an arbitrary length
line is---well, arbitrary. If the line is many wavelengths long, then
when the frequency is changed, the long lines effect kicks in and the
input Z is going to vary rapidly with respect to frequency.

Second. I believe that you need to determine the parameters of your
directional coupler. As Richard pointed out, your B1/A1 numbers are
pretty unstable.

So here's what I recommend. First verify that the "A" and "B" probes
read the same thing when connected to the same source. Then put your
pad right at the input connector of the coupler. Terminate the
reverse port and connect your VVM "A" probe to the ouput connector and
the "B" probe to the forward port.

The ratio reading is the forward coupling factor of the directional
coupler. Vary the frequency and see how this changes and note some
values. Move the "B" probe to the reverse port and terminate the
forward port. Note the readings at the same frequency.

Reverse the input and output ports and repeat the measurements.
Ideally, the data sets will track closely. If they don't then you
have a problem. Serious differences might indicate damage to the
internal terminations. This assumes that this is a true dual coupler
and not single line coupler with the termination applied to the unused
port externally.

If the numbers are consistant, then you can determine the directivity
by computing the ratio between the two readings on a given port when
the coupler is reversed.

I'm going to stop here and assume you understand the consquences of
poor directivity on measurement accuracy. If you don't then I can
expound further later.

Wes


Thanks, Wes. I'm going to have to stop the experiments for a few days, but
I'll try to get back to you.

John



John Smith October 5th 04 09:59 PM


"Wes Stewart" wrote in message
...
On Mon, 4 Oct 2004 20:58:07 -0500, "John Smith"
wrote:


Second. I believe that you need to determine the parameters of your
directional coupler. As Richard pointed out, your B1/A1 numbers are
pretty unstable.

So here's what I recommend. First verify that the "A" and "B" probes
read the same thing when connected to the same source.


I put the oscillator to the center of the HP Power Splitter. I then put a
probe tee on each of the splitter outputs followed by a 50 Ohm terminator.
The difference between the A and B channels was maybe a needle's width.
Reversing the splitter made no difference. Swapping the terminators made no
difference.


Then put your
pad right at the input connector of the coupler. Terminate the
reverse port and connect your VVM "A" probe to the ouput connector and
the "B" probe to the forward port.

The ratio reading is the forward coupling factor of the directional
coupler. Vary the frequency and see how this changes and note some
values. Move the "B" probe to the reverse port and terminate the
forward port. Note the readings at the same frequency.


From 350 MHz to 450 MHz the forward port coupling was -25.9 dB to -26.2 dB.

Reverse the input and output ports and repeat the measurements.
Ideally, the data sets will track closely.



I got the same here within about a tenth of a dB.


If they don't then you
have a problem. Serious differences might indicate damage to the
internal terminations. This assumes that this is a true dual coupler
and not single line coupler with the termination applied to the unused
port externally.

If the numbers are consistant, then you can determine the directivity
by computing the ratio between the two readings on a given port when
the coupler is reversed.

I'm going to stop here and assume you understand the consquences of
poor directivity on measurement accuracy. If you don't then I can
expound further later.

Wes



I guess it looks okay. Thanks, Wes.

John (KD5YI)



Richard Clark October 6th 04 12:08 AM

On Tue, 5 Oct 2004 12:26:06 -0500, "John Smith"
wrote:

You should get 1.0 @ 180° (the definition of a short).


Well, that's not possible when feeding a length of RG58 at 400 MHz, is it?
Remember, I said that there was about a 5 foot length of RG58 between the
directional coupler and the load. How can one get 1.0 reflected to the
coupler when the load is a short? That requires zero loss coax.


Hi John,

As Wes suggests, butt up the load against the directional coupler
output and eliminate this arbitrary loss of the 5 foot RG58. It
should also shift the readings too (you are simply walking around the
circle of constant SWR). One question that would be obviated in this
process (but I have to ask anyway) is WHERE was this short you
applied? At the output port of the coupler, or at the end of this 5
foot RG58? (Same question applies to the calibrated load).

The other measurements that you reported in response to Wes indicate
you have tracking ports (even if they are off by 4dB). As I said, it
seemed unlikely this would be a problem and it confirms the
out-of-octave specification.

73's
Richard Clark, KB7QHC

Wes Stewart October 6th 04 12:21 AM

On Tue, 5 Oct 2004 15:59:32 -0500, "John Smith"
wrote:

|
|"Wes Stewart" wrote in message
.. .
| On Mon, 4 Oct 2004 20:58:07 -0500, "John Smith"
| wrote:
|
| Second. I believe that you need to determine the parameters of your
| directional coupler. As Richard pointed out, your B1/A1 numbers are
| pretty unstable.
|
| So here's what I recommend. First verify that the "A" and "B" probes
| read the same thing when connected to the same source.
|
|I put the oscillator to the center of the HP Power Splitter. I then put a
|probe tee on each of the splitter outputs followed by a 50 Ohm terminator.
|The difference between the A and B channels was maybe a needle's width.
|Reversing the splitter made no difference. Swapping the terminators made no
|difference.

Excellent.
|
|
| Then put your
| pad right at the input connector of the coupler. Terminate the
| reverse port and connect your VVM "A" probe to the ouput connector and
| the "B" probe to the forward port.
|
| The ratio reading is the forward coupling factor of the directional
| coupler. Vary the frequency and see how this changes and note some
| values. Move the "B" probe to the reverse port and terminate the
| forward port. Note the readings at the same frequency.
|
|From 350 MHz to 450 MHz the forward port coupling was -25.9 dB to -26.2 dB.

Okay. Not per nameplate, but now you know.
|
| Reverse the input and output ports and repeat the measurements.
| Ideally, the data sets will track closely.
|
|
|I got the same here within about a tenth of a dB.

Great.
|
|
| If they don't then you
| have a problem. Serious differences might indicate damage to the
| internal terminations. This assumes that this is a true dual coupler
| and not single line coupler with the termination applied to the unused
| port externally.
|
| If the numbers are consistant, then you can determine the directivity
| by computing the ratio between the two readings on a given port when
| the coupler is reversed.

You still need to do this.

|
| I'm going to stop here and assume you understand the consquences of
| poor directivity on measurement accuracy. If you don't then I can
| expound further later.
|
| Wes
|
|
|I guess it looks okay. Thanks, Wes.

You're welcome.

Wes


John Smith October 6th 04 01:35 AM


"Wes Stewart" wrote in message
...
On Tue, 5 Oct 2004 15:59:32 -0500, "John Smith"
wrote:


| If the numbers are consistant, then you can determine the directivity
| by computing the ratio between the two readings on a given port when
| the coupler is reversed.

You still need to do this.



Do what? They are both -26.2 dB. I don't understand.

John



John Smith October 6th 04 01:35 AM


"Richard Clark" wrote in message
...
On Tue, 5 Oct 2004 12:26:06 -0500, "John Smith"
wrote:

You should get 1.0 @ 180° (the definition of a short).


Well, that's not possible when feeding a length of RG58 at 400 MHz, is it?
Remember, I said that there was about a 5 foot length of RG58 between the
directional coupler and the load. How can one get 1.0 reflected to the
coupler when the load is a short? That requires zero loss coax.


Hi John,

As Wes suggests, butt up the load against the directional coupler
output and eliminate this arbitrary loss of the 5 foot RG58. It
should also shift the readings too (you are simply walking around the
circle of constant SWR). One question that would be obviated in this
process (but I have to ask anyway) is WHERE was this short you
applied? At the output port of the coupler, or at the end of this 5
foot RG58? (Same question applies to the calibrated load).



I used a 66 inch piece of RG58 between the directional coupler and the load.
It was at the load end of this piece of coax that I calibrated with a short
and made the load measurements.


The other measurements that you reported in response to Wes indicate
you have tracking ports (even if they are off by 4dB). As I said, it
seemed unlikely this would be a problem and it confirms the
out-of-octave specification.

73's
Richard Clark, KB7QHC




John Smith October 6th 04 02:44 AM


"Wes Stewart" wrote in message
...
On Tue, 5 Oct 2004 15:59:32 -0500, "John Smith"
wrote:


| If the numbers are consistant, then you can determine the directivity
| by computing the ratio between the two readings on a given port when
| the coupler is reversed.

You still need to do this.


Hi, Wes -

I read a little about directivity (a little was all I could find). Tell me
if I measured it correctly...

Forward directivity:
Normal setup, ie coupler in the usual direction (forward). Best 50 Ohm load
I could muster on the antenna (output) terminal of the coupler. Set channel
A for 0 dB. Channel B reads -29.2 dB.

Forward Open/Short characteristic:
Remove the 50 Ohm load. Set channel A for 0 dB. Channel B reads +.4 dB. Put
the HP calibrated short on the antenna terminal. Set channel A for 0 dB.
Channel B reads -1.2 dB.

Reverse directivity:
Reverse the coupler. The antenna connector now has the oscillator applied.
The BTS terminal has the 50 Ohm load. Set channel A for 0 dB. Channel B
reads -23 dB.

Reverse Open/Short characteristic:
Remove the 50 Ohm load. Set channel A for 0 dB. Channel B reads +1.1 dB. Put
the HP calibrated short on the BTS (now the output) terminal. Set channel A
for 0 dB. Channel B reads -.5 dB.

So, in the forward direction, the directivity is -29.2 dB, and in the
reverse direction, the directivity is -23 dB. Yes?

What do I do with these results? I don't know how to apply them even though
I read the HP paper.

Thanks.

John (KD5YI)

By the way, if you (or anyone else) need to contact me via email, you can
omit the kes in the address.




All times are GMT +1. The time now is 05:49 PM.

Powered by vBulletin® Copyright ©2000 - 2025, Jelsoft Enterprises Ltd.
RadioBanter.com