![]() |
Bill Turner wrote: On 07 Dec 2003 18:25:51 GMT, (Avery Fineman) wrote: Write on the whiteboard 100 times: Inductance does not change with frequency...reactance changes with frequency. __________________________________________________ _______ Not true. Inductance and reactance are related by the formula XsubL = 2 pi F L. If XsubL has changed, then so has the inductance, and vice versa. Say what???? You have two variables that satisfy the equation: XsubL and F The equation does not mean that L varies!!!!!!!!!!! How could you possibly define it otherwise? -- Bill, W6WRT |
On Sun, 07 Dec 2003 16:14:16 -0800, Bill Turner
wrote: I use the "sub" because ascii doesn't have a lower-case L. The XsubL is merely the reactance of the inductor. If I was using Word or some other word processor I would write capital X with a subscript L. In plain English: The reactance of a coil is equal to 2 times pi times the frequency in Hz times the inductance in henries. Got it? Yes, "got it." Unfortunately it doesn't explain your hair-brained theory of variable inductance within a fixed inductor. In fact the effect you've been trying to decribe appears to be no more than an esoteric and practically-insignificant technicality - *if* indeed it exists at all. -- "I expect history will be kind to me, since I intend to write it." - Winston Churchill |
On Sun, 07 Dec 2003 16:14:16 -0800, Bill Turner
wrote: I use the "sub" because ascii doesn't have a lower-case L. The XsubL is merely the reactance of the inductor. If I was using Word or some other word processor I would write capital X with a subscript L. In plain English: The reactance of a coil is equal to 2 times pi times the frequency in Hz times the inductance in henries. Got it? Yes, "got it." Unfortunately it doesn't explain your hair-brained theory of variable inductance within a fixed inductor. In fact the effect you've been trying to decribe appears to be no more than an esoteric and practically-insignificant technicality - *if* indeed it exists at all. -- "I expect history will be kind to me, since I intend to write it." - Winston Churchill |
Bill Turner wrote:
On Sun, 07 Dec 2003 21:35:22 GMT, John Popelish wrote: You are projecting your limitations onto others. __________________________________________________ _______ I do have one limitation: I don't take insults from people I'm trying to have a discussion with. Bye. Bill, I sincerely apologize for hurting your feelings unintentionally with my clumsy comment. I should have kept strictly to inductors and away from anything that could have been interpreted as a personal attack. You may not have an impedance bridge (a limitation) but I and others do have one and they separate the components of an impedance, especially if you take two or more readings at different frequencies and solve a bit of math. -- John Popelish |
Bill Turner wrote:
On Sun, 07 Dec 2003 21:35:22 GMT, John Popelish wrote: You are projecting your limitations onto others. __________________________________________________ _______ I do have one limitation: I don't take insults from people I'm trying to have a discussion with. Bye. Bill, I sincerely apologize for hurting your feelings unintentionally with my clumsy comment. I should have kept strictly to inductors and away from anything that could have been interpreted as a personal attack. You may not have an impedance bridge (a limitation) but I and others do have one and they separate the components of an impedance, especially if you take two or more readings at different frequencies and solve a bit of math. -- John Popelish |
I read in sci.electronics.design that Paul Keinanen
wrote (in ) about 'Winding coils', on Sun, 7 Dec 2003: On Sun, 7 Dec 2003 19:13:36 +0000, John Woodgate wrote: Low-frequency iron-cored coils are quite another matter; the inductance varies with frequency, voltage, temperature, previous history and the state of the tide on Europa. I assume that you are referring to DC biased iron cores (without an air gap) or some high permeability ferrites with a strong DC bias current. These do indeed show a variation of inductance depending on the DC bias current. Not only that, the inductance can vary with the AC voltage applied, most notably when saturation is approached, but it can also happen with silicon iron at very low inductions. Nickel-iron alloys don't normally show this 'bottom bend' effect. -- Regards, John Woodgate, OOO - Own Opinions Only. http://www.jmwa.demon.co.uk Interested in professional sound reinforcement and distribution? Then go to http://www.isce.org.uk PLEASE do NOT copy news posts to me by E-MAIL! |
I read in sci.electronics.design that Paul Keinanen
wrote (in ) about 'Winding coils', on Sun, 7 Dec 2003: On Sun, 7 Dec 2003 19:13:36 +0000, John Woodgate wrote: Low-frequency iron-cored coils are quite another matter; the inductance varies with frequency, voltage, temperature, previous history and the state of the tide on Europa. I assume that you are referring to DC biased iron cores (without an air gap) or some high permeability ferrites with a strong DC bias current. These do indeed show a variation of inductance depending on the DC bias current. Not only that, the inductance can vary with the AC voltage applied, most notably when saturation is approached, but it can also happen with silicon iron at very low inductions. Nickel-iron alloys don't normally show this 'bottom bend' effect. -- Regards, John Woodgate, OOO - Own Opinions Only. http://www.jmwa.demon.co.uk Interested in professional sound reinforcement and distribution? Then go to http://www.isce.org.uk PLEASE do NOT copy news posts to me by E-MAIL! |
In article , Bill Turner
writes: On 07 Dec 2003 18:25:51 GMT, (Avery Fineman) wrote: Write on the whiteboard 100 times: Inductance does not change with frequency...reactance changes with frequency. _________________________________________________ ________ Not true. Inductance and reactance are related by the formula XsubL = 2 pi F L. If XsubL has changed, then so has the inductance, and vice versa. How could you possibly define it otherwise? Bill, I can get down to first principles if necessary, but that isn't necessary, is it? INDUCTANCE doesn't change over frequency...even above the "self-resonance" due to distributed capacity between windings. That's very basic and applies up into the region where the frequency is so high the whole "coil" structure starts behaving like a distributed-constant conglomeration of equivalent parts. But, that's a specialty area and far above any practical application of home-made coils for RF. Reactance is a function of frequency and inductance. The reactance of an inductor DOES change over frequency. That's also very basic. For _practical_ home-made coils, the only major concern is the distributed capacity of the coil structure. Distributed capacity is the _equivalent_ of a fixed, parallel capacitor across the pure inductor part of the coil. That L and C will determine the "self resonance" of the structure. To find the distributed capacity of an inductor (the equivalent of a fixed parallel capacity connected across the inductor), the method described in the "Reference Data for Radio Engineers," fourth edition, 1956, ITT (aka "Green Bible"), chapter 10, pp 268- 269 can be used as follows: Using a Q Meter or other instrument with a calibrated variable capacitor, resonate the parallel L-C with the capacitor at two frequencies exactly an octave apart (1:2 ratio). Take the difference of the two variable capacitor resonating values as "deltac." Let "freqsq" be the _square_ of the highest of the two frequencies used. For uHy, pFd, and MHz: L = (19,000) / (freqsq x deltac) Inductance L is the "true" inductance of the coil, separated from the distributed capacity. The constant of "19,000" is a simple approximation considering that 1956 was the age of slide rules and electromechanical four- function calculators. If the parallel resonating capacitor is well- calibrated, the "true inductance" formula works out well. If the parallel resonating capacitor is not calibrated, forget the whole thing; there are several C-meters on the market that can allow rather precise +/- 0.1 pFd resolution calibration if anyone is into home metrology. Anyone wishing to play with simple algebra can figure out the formula from basic resonance equation at two frequencies exactly an octave apart. That will result in the true mathematical value of the constant given in the Green Bible. :-) Len Anderson retired (from regular hours) electronic engineer person |
In article , Bill Turner
writes: On 07 Dec 2003 18:25:51 GMT, (Avery Fineman) wrote: Write on the whiteboard 100 times: Inductance does not change with frequency...reactance changes with frequency. _________________________________________________ ________ Not true. Inductance and reactance are related by the formula XsubL = 2 pi F L. If XsubL has changed, then so has the inductance, and vice versa. How could you possibly define it otherwise? Bill, I can get down to first principles if necessary, but that isn't necessary, is it? INDUCTANCE doesn't change over frequency...even above the "self-resonance" due to distributed capacity between windings. That's very basic and applies up into the region where the frequency is so high the whole "coil" structure starts behaving like a distributed-constant conglomeration of equivalent parts. But, that's a specialty area and far above any practical application of home-made coils for RF. Reactance is a function of frequency and inductance. The reactance of an inductor DOES change over frequency. That's also very basic. For _practical_ home-made coils, the only major concern is the distributed capacity of the coil structure. Distributed capacity is the _equivalent_ of a fixed, parallel capacitor across the pure inductor part of the coil. That L and C will determine the "self resonance" of the structure. To find the distributed capacity of an inductor (the equivalent of a fixed parallel capacity connected across the inductor), the method described in the "Reference Data for Radio Engineers," fourth edition, 1956, ITT (aka "Green Bible"), chapter 10, pp 268- 269 can be used as follows: Using a Q Meter or other instrument with a calibrated variable capacitor, resonate the parallel L-C with the capacitor at two frequencies exactly an octave apart (1:2 ratio). Take the difference of the two variable capacitor resonating values as "deltac." Let "freqsq" be the _square_ of the highest of the two frequencies used. For uHy, pFd, and MHz: L = (19,000) / (freqsq x deltac) Inductance L is the "true" inductance of the coil, separated from the distributed capacity. The constant of "19,000" is a simple approximation considering that 1956 was the age of slide rules and electromechanical four- function calculators. If the parallel resonating capacitor is well- calibrated, the "true inductance" formula works out well. If the parallel resonating capacitor is not calibrated, forget the whole thing; there are several C-meters on the market that can allow rather precise +/- 0.1 pFd resolution calibration if anyone is into home metrology. Anyone wishing to play with simple algebra can figure out the formula from basic resonance equation at two frequencies exactly an octave apart. That will result in the true mathematical value of the constant given in the Green Bible. :-) Len Anderson retired (from regular hours) electronic engineer person |
Bill Turner wrote: On 08 Dec 2003 20:09:43 GMT, (Avery Fineman) wrote: INDUCTANCE doesn't change over frequency __________________________________________________ _______ I maintain it does. Otherwise the formula X=2piFL is invalid. NO! In the above equation, X varies when F varies. The equation does NOT mean that L varies as F varies. Is that what you're saying? I understand what you're saying about the inductance of a coil being fixed and the reactance is the net result of that fixed inductance plus the effect of the parasitic capacitance between windings, vs frequency, of course. If one chooses to *model* a coil that way, I have no objection. You will no doubt arrive at the correct reactance for a given frequency. The disagreement here seems to depend on how one defines what inductance is. I maintain that inductance of a coil is nothing more than the reactance divided by 2piF, as derived from the formula above. Do you disagree with that? That formula has been taught for decades. Are you saying it is wrong? The formula is fine. Your understanding of it is wrong. X is inductive reactance. F is frequency. L is inductance. F is a variable, L is fixed and X (the reactance of L at the frequency) varies as the frequency varies. -- Bill, W6WRT |
Bill Turner wrote: On 08 Dec 2003 20:09:43 GMT, (Avery Fineman) wrote: INDUCTANCE doesn't change over frequency __________________________________________________ _______ I maintain it does. Otherwise the formula X=2piFL is invalid. NO! In the above equation, X varies when F varies. The equation does NOT mean that L varies as F varies. Is that what you're saying? I understand what you're saying about the inductance of a coil being fixed and the reactance is the net result of that fixed inductance plus the effect of the parasitic capacitance between windings, vs frequency, of course. If one chooses to *model* a coil that way, I have no objection. You will no doubt arrive at the correct reactance for a given frequency. The disagreement here seems to depend on how one defines what inductance is. I maintain that inductance of a coil is nothing more than the reactance divided by 2piF, as derived from the formula above. Do you disagree with that? That formula has been taught for decades. Are you saying it is wrong? The formula is fine. Your understanding of it is wrong. X is inductive reactance. F is frequency. L is inductance. F is a variable, L is fixed and X (the reactance of L at the frequency) varies as the frequency varies. -- Bill, W6WRT |
On Mon, 08 Dec 2003 19:46:43 -0800, Bill Turner
wrote: On 08 Dec 2003 20:09:43 GMT, (Avery Fineman) wrote: INDUCTANCE doesn't change over frequency _________________________________________________ ________ I maintain it does. Otherwise the formula X=2piFL is invalid. Is that what you're saying? I understand what you're saying about the inductance of a coil being fixed and the reactance is the net result of that fixed inductance plus the effect of the parasitic capacitance between windings, vs frequency, of course. If one chooses to *model* a coil that way, I have no objection. You will no doubt arrive at the correct reactance for a given frequency. That is what everybody is trying to say. It is like discussing is a candle _emitting_light_ into the room or is the candle _absorbing_darkness_. This becomes quite apparent when the wick of the candle is black when the candle has been put out, clearly it has absorbed a lot of darkness :-). In an incandescent lamp, the electric current will constantly renew the filament, thus preventing a lot of darkness being concentrated on the filament. One could develop quite scientific methods to measure the amount of darkness absorbed and predict the behaviour of other lamps. This can also be debated successfully for a quite a while, until some serious disagreeing measurements are brought into the discussion. The disagreement here seems to depend on how one defines what inductance is. It has often been defined by the ability to store energy. I maintain that inductance of a coil is nothing more than the reactance divided by 2piF, as derived from the formula above. L = Xl / (2 pi f) applies only to _pure_inductive Xl It does _not_ apply to L = X / (2 pi f) in which X is some combination of Xl and Xc ! Do you disagree with that? That formula has been taught for decades. Are you saying it is wrong? The formulas taught for decades a Xl = 2 pi f L _and _ Xc = -1/(2 pi f C) How do you arrive to the incorrect L = X / (2 pi f) from the equations above ? Please note, it is the magnitude of X what you are measuring with some simple test gear, not Xl. Thus, the original claim is an artefact of the measurement method. Paul OH3LWR |
On Mon, 08 Dec 2003 19:46:43 -0800, Bill Turner
wrote: On 08 Dec 2003 20:09:43 GMT, (Avery Fineman) wrote: INDUCTANCE doesn't change over frequency _________________________________________________ ________ I maintain it does. Otherwise the formula X=2piFL is invalid. Is that what you're saying? I understand what you're saying about the inductance of a coil being fixed and the reactance is the net result of that fixed inductance plus the effect of the parasitic capacitance between windings, vs frequency, of course. If one chooses to *model* a coil that way, I have no objection. You will no doubt arrive at the correct reactance for a given frequency. That is what everybody is trying to say. It is like discussing is a candle _emitting_light_ into the room or is the candle _absorbing_darkness_. This becomes quite apparent when the wick of the candle is black when the candle has been put out, clearly it has absorbed a lot of darkness :-). In an incandescent lamp, the electric current will constantly renew the filament, thus preventing a lot of darkness being concentrated on the filament. One could develop quite scientific methods to measure the amount of darkness absorbed and predict the behaviour of other lamps. This can also be debated successfully for a quite a while, until some serious disagreeing measurements are brought into the discussion. The disagreement here seems to depend on how one defines what inductance is. It has often been defined by the ability to store energy. I maintain that inductance of a coil is nothing more than the reactance divided by 2piF, as derived from the formula above. L = Xl / (2 pi f) applies only to _pure_inductive Xl It does _not_ apply to L = X / (2 pi f) in which X is some combination of Xl and Xc ! Do you disagree with that? That formula has been taught for decades. Are you saying it is wrong? The formulas taught for decades a Xl = 2 pi f L _and _ Xc = -1/(2 pi f C) How do you arrive to the incorrect L = X / (2 pi f) from the equations above ? Please note, it is the magnitude of X what you are measuring with some simple test gear, not Xl. Thus, the original claim is an artefact of the measurement method. Paul OH3LWR |
On Tue, 09 Dec 2003 04:59:29 GMT, wrote:
Bill Turner wrote: On 08 Dec 2003 20:09:43 GMT, (Avery Fineman) wrote: INDUCTANCE doesn't change over frequency __________________________________________________ _______ I maintain it does. Otherwise the formula X=2piFL is invalid. NO! In the above equation, X varies when F varies. The equation does NOT mean that L varies as F varies. That's right. You end up with X/2piF and X and F are inter-related; they are not independent variables. -- "I expect history will be kind to me, since I intend to write it." - Winston Churchill |
1) you guys are just arguing symantics. You both know what really happens.
".... "Inductance" vs. the total reactance measuring as inductive...." call it what you like. 1a) You are also both using (some might say mis-using) the term "linear" to mean "varies linearly with..." rather than the more common meaning that superposition applies. RLC sircuits are linear. Any given parameter may not vary linearly as the frequency is varied. Also, this use of 'linear' depends upon the type of scale being used--log or linear. 2) John, You better re-think your last statement about the series equivalent of a practical coil. It implies that there is some way to measure a low Z at the resonance of the coil under discussion. You say: "The series equivalent [impedance ? Steve] goes down as the frequency increases, and goes to zero at resonance. " While a series resonant LC exhibits this behavior, the series equivalent of a practical coil does not do this. The series equivalent must do the same thing as the parallel equivalent -- namely go to a high impedance at resonance. That's why it is called *equivalent*--the total, or terminal impedance is equal for the two representations (at a single frequency). Pretty sure I got that right.... Steve k]9]d]c]i A practical coil usually goes to parallel resonance - at which the series equivalent does not go to zero "John Woodgate" wrote in message ... I read in sci.electronics.design that Bill Turner wrote (in ) about 'Winding coils', on Sun, 7 Dec 2003: Both statements are true and easily provable. A simple air core coil which measures one microhenry at a low frequency may have an inductance of several millihenries (or even henries) when near its self resonant frequency. This is what happens to the *parallel equivalent* inductance. The series equivalent goes down as the frequency increases, and goes to zero at resonance. -- Regards, John Woodgate, OOO - Own Opinions Only. http://www.jmwa.demon.co.uk Interested in professional sound reinforcement and distribution? Then go to http://www.isce.org.uk PLEASE do NOT copy news posts to me by E-MAIL! |
1) you guys are just arguing symantics. You both know what really happens.
".... "Inductance" vs. the total reactance measuring as inductive...." call it what you like. 1a) You are also both using (some might say mis-using) the term "linear" to mean "varies linearly with..." rather than the more common meaning that superposition applies. RLC sircuits are linear. Any given parameter may not vary linearly as the frequency is varied. Also, this use of 'linear' depends upon the type of scale being used--log or linear. 2) John, You better re-think your last statement about the series equivalent of a practical coil. It implies that there is some way to measure a low Z at the resonance of the coil under discussion. You say: "The series equivalent [impedance ? Steve] goes down as the frequency increases, and goes to zero at resonance. " While a series resonant LC exhibits this behavior, the series equivalent of a practical coil does not do this. The series equivalent must do the same thing as the parallel equivalent -- namely go to a high impedance at resonance. That's why it is called *equivalent*--the total, or terminal impedance is equal for the two representations (at a single frequency). Pretty sure I got that right.... Steve k]9]d]c]i A practical coil usually goes to parallel resonance - at which the series equivalent does not go to zero "John Woodgate" wrote in message ... I read in sci.electronics.design that Bill Turner wrote (in ) about 'Winding coils', on Sun, 7 Dec 2003: Both statements are true and easily provable. A simple air core coil which measures one microhenry at a low frequency may have an inductance of several millihenries (or even henries) when near its self resonant frequency. This is what happens to the *parallel equivalent* inductance. The series equivalent goes down as the frequency increases, and goes to zero at resonance. -- Regards, John Woodgate, OOO - Own Opinions Only. http://www.jmwa.demon.co.uk Interested in professional sound reinforcement and distribution? Then go to http://www.isce.org.uk PLEASE do NOT copy news posts to me by E-MAIL! |
Gents,
Another practical consideration. Another area where caution is advised--paralleling bypass caps. In solid state Power Amplifier design, such a configuratin can cause problems because there is a point where one is above self resonance and acts like an inductance in parallel with the other cap which is still capacitive-- thus, resonance and no bypass. Been there, done that. We put a small Z in between. Frequently a small bead or resistor if possible. Seems there is an equivalent problem with series inductors. -- Steve N, K,9 d, c. i "Bill Turner" wrote in message ... On Sun, 7 Dec 2003 19:17:08 +0000, John Woodgate wrote: This is a 1920s problem. Just as you parallel capacitors of different type, electrolytic, metallized foil and ceramic, to get a wideband component, so you put inductors of different construction in series to get a wide band component. You can wind them all on a bit of wax- impregnated dowel if you like. (;-) __________________________________________________ _______ That will work, no doubt. My point was that it takes some serious engineering and careful testing; you can't just wrap some wire on a form and expect it to work correctly across a wide range of frequencies. -- Bill, W6WRT |
Gents,
Another practical consideration. Another area where caution is advised--paralleling bypass caps. In solid state Power Amplifier design, such a configuratin can cause problems because there is a point where one is above self resonance and acts like an inductance in parallel with the other cap which is still capacitive-- thus, resonance and no bypass. Been there, done that. We put a small Z in between. Frequently a small bead or resistor if possible. Seems there is an equivalent problem with series inductors. -- Steve N, K,9 d, c. i "Bill Turner" wrote in message ... On Sun, 7 Dec 2003 19:17:08 +0000, John Woodgate wrote: This is a 1920s problem. Just as you parallel capacitors of different type, electrolytic, metallized foil and ceramic, to get a wideband component, so you put inductors of different construction in series to get a wide band component. You can wind them all on a bit of wax- impregnated dowel if you like. (;-) __________________________________________________ _______ That will work, no doubt. My point was that it takes some serious engineering and careful testing; you can't just wrap some wire on a form and expect it to work correctly across a wide range of frequencies. -- Bill, W6WRT |
You're still doing it. Paul (I think) said "measure" and Bill, no, looks
like Len (I think) said "finding", meaning "calculating". "Avery Fineman" wrote in message ... In article , Bill Turner writes: On Sun, 07 Dec 2003 13:55:35 +0200, Paul Keinanen wrote: .snip Not only can you *not* measure them separately, they can not be physically separated either, since the parasitic capacitance is always present between adjacent windings.... Nonsense. General Radio had a nice little formula way back before 1956 for finding the distributed capacity of an inductor. Len Anderson retired (from regular hours) electronic engineer person |
You're still doing it. Paul (I think) said "measure" and Bill, no, looks
like Len (I think) said "finding", meaning "calculating". "Avery Fineman" wrote in message ... In article , Bill Turner writes: On Sun, 07 Dec 2003 13:55:35 +0200, Paul Keinanen wrote: .snip Not only can you *not* measure them separately, they can not be physically separated either, since the parasitic capacitance is always present between adjacent windings.... Nonsense. General Radio had a nice little formula way back before 1956 for finding the distributed capacity of an inductor. Len Anderson retired (from regular hours) electronic engineer person |
OOPS Bill. By the formula, change F and Xl changes! Avery said below;
"reactance changes with frequency" and is correct. Also, if you change L then Xl changes, but that is not what he said. -- Steve N, K,9 d, c. i "Bill Turner" wrote in message ... On 07 Dec 2003 18:25:51 GMT, (Avery Fineman) wrote: Write on the whiteboard 100 times: Inductance does not change with frequency...reactance changes with frequency. __________________________________________________ _______ Not true. Inductance and reactance are related by the formula XsubL = 2 pi F L. If XsubL has changed, then so has the inductance, and vice versa. How could you possibly define it otherwise? -- Bill, W6WRT |
OOPS Bill. By the formula, change F and Xl changes! Avery said below;
"reactance changes with frequency" and is correct. Also, if you change L then Xl changes, but that is not what he said. -- Steve N, K,9 d, c. i "Bill Turner" wrote in message ... On 07 Dec 2003 18:25:51 GMT, (Avery Fineman) wrote: Write on the whiteboard 100 times: Inductance does not change with frequency...reactance changes with frequency. __________________________________________________ _______ Not true. Inductance and reactance are related by the formula XsubL = 2 pi F L. If XsubL has changed, then so has the inductance, and vice versa. How could you possibly define it otherwise? -- Bill, W6WRT |
Ahhhh! So there does seem to be a mis interpretation of the formula
here.-Steve Bill, I believe you are placing the dependant and independent variables in the wrong place. The Xl is the dependant variable. Xl *depends upon* F and L, not the other way around. That is, given an L and F you calculate the X. X is the answer. Only if you know there are no other contributing factors can the formula be used the other way, because it does not factor them (the parasitic capacitance) in. That is, Given an X measurement you can not tell the inductance, only the *equivalent total* inductive reactance. -- Steve N, K,9 d, c. i "Bill Turner" wrote in message ... On 08 Dec 2003 20:09:43 GMT, (Avery Fineman) wrote: INDUCTANCE doesn't change over frequency __________________________________________________ _______ I maintain it does. Otherwise the formula X=2piFL is invalid. Is that what you're saying? I understand what you're saying about the inductance of a coil being fixed and the reactance is the net result of that fixed inductance plus the effect of the parasitic capacitance between windings, vs frequency, of course. If one chooses to *model* a coil that way, I have no objection. You will no doubt arrive at the correct reactance for a given frequency. The disagreement here seems to depend on how one defines what inductance is. I maintain that inductance of a coil is nothing more than the reactance divided by 2piF, as derived from the formula above. Do you disagree with that? That formula has been taught for decades. Are you saying it is wrong? -- Bill, W6WRT |
Ahhhh! So there does seem to be a mis interpretation of the formula
here.-Steve Bill, I believe you are placing the dependant and independent variables in the wrong place. The Xl is the dependant variable. Xl *depends upon* F and L, not the other way around. That is, given an L and F you calculate the X. X is the answer. Only if you know there are no other contributing factors can the formula be used the other way, because it does not factor them (the parasitic capacitance) in. That is, Given an X measurement you can not tell the inductance, only the *equivalent total* inductive reactance. -- Steve N, K,9 d, c. i "Bill Turner" wrote in message ... On 08 Dec 2003 20:09:43 GMT, (Avery Fineman) wrote: INDUCTANCE doesn't change over frequency __________________________________________________ _______ I maintain it does. Otherwise the formula X=2piFL is invalid. Is that what you're saying? I understand what you're saying about the inductance of a coil being fixed and the reactance is the net result of that fixed inductance plus the effect of the parasitic capacitance between windings, vs frequency, of course. If one chooses to *model* a coil that way, I have no objection. You will no doubt arrive at the correct reactance for a given frequency. The disagreement here seems to depend on how one defines what inductance is. I maintain that inductance of a coil is nothing more than the reactance divided by 2piF, as derived from the formula above. Do you disagree with that? That formula has been taught for decades. Are you saying it is wrong? -- Bill, W6WRT |
There'something sour here. Way down ...
"Paul Keinanen" wrote in message ... [snip] The problem with circuits containing both inductances and capacitances is that in one kind of reactance, there is a +90 degree phase shift and the other with -90 degree phase shift. Thus, when these are combined, they partially cancel each other, producing different magnitudes and some phase shift between -90 and +90 degrees. If only the resultant magnitude is used (and the resultant phase is ignored), this would give the false impression that the inductance changes with frequency. I don't quite follow where you are going here. below the self resonant freq the angle will be +90 (minus a little for what ever resistance is there). The rest of this about measuring the energy from DC, I don't think is at all practical. [snip] the inductance could be measured in a different way. ... the energy stored in the inductance is W = I*I*L/2. ... ...cut the DC current...dissipate the energy in some kind of integrating load across L. Even if there is a significant capacitance[snip] ...the energy would bounce back between L and C, but finally it would be dissipated by the external load. ... Thus using this measurement method, the value of L would be the same regardless if C is present or not.... Paul OH3LWR OK. so then, how do you propose to measure this energy? I don't think it is practical. -- Steve N, K,9 d, c. i |
There'something sour here. Way down ...
"Paul Keinanen" wrote in message ... [snip] The problem with circuits containing both inductances and capacitances is that in one kind of reactance, there is a +90 degree phase shift and the other with -90 degree phase shift. Thus, when these are combined, they partially cancel each other, producing different magnitudes and some phase shift between -90 and +90 degrees. If only the resultant magnitude is used (and the resultant phase is ignored), this would give the false impression that the inductance changes with frequency. I don't quite follow where you are going here. below the self resonant freq the angle will be +90 (minus a little for what ever resistance is there). The rest of this about measuring the energy from DC, I don't think is at all practical. [snip] the inductance could be measured in a different way. ... the energy stored in the inductance is W = I*I*L/2. ... ...cut the DC current...dissipate the energy in some kind of integrating load across L. Even if there is a significant capacitance[snip] ...the energy would bounce back between L and C, but finally it would be dissipated by the external load. ... Thus using this measurement method, the value of L would be the same regardless if C is present or not.... Paul OH3LWR OK. so then, how do you propose to measure this energy? I don't think it is practical. -- Steve N, K,9 d, c. i |
On Tue, 09 Dec 2003 09:19:23 -0800, Bill Turner
wrote: Perhaps an example will make it clear. Suppose you have a coil which measures 1 uH at 1 MHz. It is known to have a self-resonant (parallel) frequency of 100 MHz. You measure its reactance at 1 MHz using the formula X=2*pi*F and find it to be 6.28 ohms. At 2 MHz you find it to be 12.56 ohms. At 10 MHz you find it to be 62.8 ohms. So far the reactance is changing linearly with respect to frequency. (Actually it is not perfectly linear, but the difference at these frequencies is small and probably would not be observed with run of the mill test equipment.) But, as you approach 100 MHz, you find the change is obviously no longer linear. At 95 MHz you would expect the reactance to be 6.28*95=596.6 ohms, but much to your surprise, it measures 1000 ohms. At 99 MHz, instead of the expected 6.28*99=621.72 ohms, it measures 50,000 ohms!! All the above is perfectly normal and easily observable. My point is that when a coil measures 50,000 ohms at 99 MHz, its inductance HAS TO BE L=X/(2*pi*F), or 50,000/(6.28*99)=80.4 uH! This is not an illusion. If you have an inductance meter which uses 99 MHz as a test frequency, it WILL MEASURE 80.4 uH. And therefore, I maintain that inductance DOES vary with frequency. How can it be otherwise? Measurement errors? I don't know enough about your way of working to say. But thanks for giving your worked example. It'll no doubt help to pin down the exact area of disagreement between us. -- "I expect history will be kind to me, since I intend to write it." - Winston Churchill |
On Tue, 09 Dec 2003 09:19:23 -0800, Bill Turner
wrote: Perhaps an example will make it clear. Suppose you have a coil which measures 1 uH at 1 MHz. It is known to have a self-resonant (parallel) frequency of 100 MHz. You measure its reactance at 1 MHz using the formula X=2*pi*F and find it to be 6.28 ohms. At 2 MHz you find it to be 12.56 ohms. At 10 MHz you find it to be 62.8 ohms. So far the reactance is changing linearly with respect to frequency. (Actually it is not perfectly linear, but the difference at these frequencies is small and probably would not be observed with run of the mill test equipment.) But, as you approach 100 MHz, you find the change is obviously no longer linear. At 95 MHz you would expect the reactance to be 6.28*95=596.6 ohms, but much to your surprise, it measures 1000 ohms. At 99 MHz, instead of the expected 6.28*99=621.72 ohms, it measures 50,000 ohms!! All the above is perfectly normal and easily observable. My point is that when a coil measures 50,000 ohms at 99 MHz, its inductance HAS TO BE L=X/(2*pi*F), or 50,000/(6.28*99)=80.4 uH! This is not an illusion. If you have an inductance meter which uses 99 MHz as a test frequency, it WILL MEASURE 80.4 uH. And therefore, I maintain that inductance DOES vary with frequency. How can it be otherwise? Measurement errors? I don't know enough about your way of working to say. But thanks for giving your worked example. It'll no doubt help to pin down the exact area of disagreement between us. -- "I expect history will be kind to me, since I intend to write it." - Winston Churchill |
Perhaps an example will make it clear.
Suppose you have a coil which measures 1 uH at 1 MHz. It is known to have a self-resonant (parallel) frequency of 100 MHz. You measure its reactance at 1 MHz using the formula X=2*pi*F and find it to be 6.28 ohms. At 2 MHz you find it to be 12.56 ohms. At 10 MHz you find it to be 62.8 ohms. So far the reactance is changing linearly with respect to frequency. (Actually it is not perfectly linear, but the difference at these frequencies is small and probably would not be observed with run of the mill test equipment.) But, as you approach 100 MHz, you find the change is obviously no longer linear. At 95 MHz you would expect the reactance to be 6.28*95=596.6 ohms, but much to your surprise, it measures 1000 ohms. At 99 MHz, instead of the expected 6.28*99=621.72 ohms, it measures 50,000 ohms!! All the above is perfectly normal and easily observable. My point is that when a coil measures 50,000 ohms at 99 MHz, its inductance HAS TO BE L=X/(2*pi*F), or 50,000/(6.28*99)=80.4 uH! This is not an illusion. If you have an inductance meter which uses 99 MHz as a test frequency, it WILL MEASURE 80.4 uH. And therefore, I maintain that inductance DOES vary with frequency. How can it be otherwise? -- Bill, W6WRT The inductance is not changing. What you are measuring is not pure inductance but the coil has a stray capacitance. That is what is making the coil seof resonate. YOu did not say what hapens at 110 mhz, 200 mhz, and 500 mhz, if you did , it would measuer capacitance reactance. How do you change a coil into a capacitor ? You don't , but the effect of reactance has. Look at it this from a totally differant angle. You stick the leads of a DV voltmeter in the wall socket. It does not show any deflection other than maybe the first jump when it is plugged in. Does that mean there is no voltage or power in the circuit, I think not. Stick your fingers in it and see what hapens :-) Your method is flawed in the same way, you only measured inductance ( not really that , but the inductive reactance at a given frequency, but did not measuer capcitance. Where did the capacitance come from ? It is what makes the coil selfresonante. If you measuer a circuit that has inductance, capacitance and resistance, depending on if it is series or pareallel resonate here is what will hapen. As the frequency is increaced the inductance reactance will increace, it will measuer resistance at the reosnant frequency , then a large capacitance reactance and then a small capacitance reactance or else the reverse will hapen, capacitive reactance, resisstance, inductive reactance. However none of the actual inductance, capacitance or resistance values will change. YOu are confusing inductacne and reactance. YOu are only seeing one part of the big picture. YOu have to look at several formulars to see what is going on in a circuit that has inductance and capacitance. |
Perhaps an example will make it clear.
Suppose you have a coil which measures 1 uH at 1 MHz. It is known to have a self-resonant (parallel) frequency of 100 MHz. You measure its reactance at 1 MHz using the formula X=2*pi*F and find it to be 6.28 ohms. At 2 MHz you find it to be 12.56 ohms. At 10 MHz you find it to be 62.8 ohms. So far the reactance is changing linearly with respect to frequency. (Actually it is not perfectly linear, but the difference at these frequencies is small and probably would not be observed with run of the mill test equipment.) But, as you approach 100 MHz, you find the change is obviously no longer linear. At 95 MHz you would expect the reactance to be 6.28*95=596.6 ohms, but much to your surprise, it measures 1000 ohms. At 99 MHz, instead of the expected 6.28*99=621.72 ohms, it measures 50,000 ohms!! All the above is perfectly normal and easily observable. My point is that when a coil measures 50,000 ohms at 99 MHz, its inductance HAS TO BE L=X/(2*pi*F), or 50,000/(6.28*99)=80.4 uH! This is not an illusion. If you have an inductance meter which uses 99 MHz as a test frequency, it WILL MEASURE 80.4 uH. And therefore, I maintain that inductance DOES vary with frequency. How can it be otherwise? -- Bill, W6WRT The inductance is not changing. What you are measuring is not pure inductance but the coil has a stray capacitance. That is what is making the coil seof resonate. YOu did not say what hapens at 110 mhz, 200 mhz, and 500 mhz, if you did , it would measuer capacitance reactance. How do you change a coil into a capacitor ? You don't , but the effect of reactance has. Look at it this from a totally differant angle. You stick the leads of a DV voltmeter in the wall socket. It does not show any deflection other than maybe the first jump when it is plugged in. Does that mean there is no voltage or power in the circuit, I think not. Stick your fingers in it and see what hapens :-) Your method is flawed in the same way, you only measured inductance ( not really that , but the inductive reactance at a given frequency, but did not measuer capcitance. Where did the capacitance come from ? It is what makes the coil selfresonante. If you measuer a circuit that has inductance, capacitance and resistance, depending on if it is series or pareallel resonate here is what will hapen. As the frequency is increaced the inductance reactance will increace, it will measuer resistance at the reosnant frequency , then a large capacitance reactance and then a small capacitance reactance or else the reverse will hapen, capacitive reactance, resisstance, inductive reactance. However none of the actual inductance, capacitance or resistance values will change. YOu are confusing inductacne and reactance. YOu are only seeing one part of the big picture. YOu have to look at several formulars to see what is going on in a circuit that has inductance and capacitance. |
In article , Bill Turner
writes: The disagreement here seems to depend on how one defines what inductance is. I maintain that inductance of a coil is nothing more than the reactance divided by 2piF, as derived from the formula above. Do you disagree with that? That formula has been taught for decades. Are you saying it is wrong? I'm saying that the student doesn't understand inductance. Inductance does NOT vary over frequency for any coil of wire under its self-resonance. Reactance varies over frequency with inductance fixed...directly proportional to frequency. Inductance doesn't vary. Yes, you can FIND inductance in Henries if you measure its reactance at a particular frequency. Inductance in Henries has NOT changed by doing so. Inductance in Henries remains constant. [feel free to quibble over the spelling of "Henries" v. "Henrys" :-) ] If your reactance-measuring gizmo is not calibrated properly, then its readings will show an APPARENT change in inductance. The inductance still hasn't changed...only the calibration of the gizmo is off. Don't get all wound up and take a turn for the worse... Len Anderson retired (from regular hours) electronic engineer person |
In article , Bill Turner
writes: The disagreement here seems to depend on how one defines what inductance is. I maintain that inductance of a coil is nothing more than the reactance divided by 2piF, as derived from the formula above. Do you disagree with that? That formula has been taught for decades. Are you saying it is wrong? I'm saying that the student doesn't understand inductance. Inductance does NOT vary over frequency for any coil of wire under its self-resonance. Reactance varies over frequency with inductance fixed...directly proportional to frequency. Inductance doesn't vary. Yes, you can FIND inductance in Henries if you measure its reactance at a particular frequency. Inductance in Henries has NOT changed by doing so. Inductance in Henries remains constant. [feel free to quibble over the spelling of "Henries" v. "Henrys" :-) ] If your reactance-measuring gizmo is not calibrated properly, then its readings will show an APPARENT change in inductance. The inductance still hasn't changed...only the calibration of the gizmo is off. Don't get all wound up and take a turn for the worse... Len Anderson retired (from regular hours) electronic engineer person |
Bill Turner wrote: On Tue, 09 Dec 2003 04:59:29 GMT, wrote: I maintain it does. Otherwise the formula X=2piFL is invalid. NO! In the above equation, X varies when F varies. The equation does NOT mean that L varies as F varies. __________________________________________________ _______ Perhaps an example will make it clear. Suppose you have a coil which measures 1 uH at 1 MHz. It is known to have a self-resonant (parallel) frequency of 100 MHz. You measure its reactance at 1 MHz using the formula X=2*pi*F and find it to be 6.28 ohms. At 2 MHz you find it to be 12.56 ohms. At 10 MHz you find it to be 62.8 ohms. So far the reactance is changing linearly with respect to frequency. (Actually it is not perfectly linear, but the difference at these frequencies is small and probably would not be observed with run of the mill test equipment.) But, as you approach 100 MHz, you find the change is obviously no longer linear. At 95 MHz you would expect the reactance to be 6.28*95=596.6 ohms, but much to your surprise, it measures 1000 ohms. At 99 MHz, instead of the expected 6.28*99=621.72 ohms, it measures 50,000 ohms!! All the above is perfectly normal and easily observable. My point is that when a coil measures 50,000 ohms at 99 MHz, its inductance HAS TO BE L=X/(2*pi*F), or 50,000/(6.28*99)=80.4 uH! Here are your own words: "At that self-resonant frequency, the coil is behaving like a parallel resonant circuit, which of course it is, due to the parasitic capacitance between each winding." Your example ignores the capacitance, which you have stated exists. There is nothing in your formula that addresses it. You cannot use the formula or the math above (in your post) to support your point of view, because it does not contain any term for capacitance. The capacitance exists, and exhibits a larger and larger affect on the circuit as the frequency increases from 1 - 99 mHz. This is not an illusion. If you have an inductance meter which uses 99 MHz as a test frequency, it WILL MEASURE 80.4 uH. And therefore, I maintain that inductance DOES vary with frequency. How can it be otherwise? -- Bill, W6WRT |
Bill Turner wrote: On Tue, 09 Dec 2003 04:59:29 GMT, wrote: I maintain it does. Otherwise the formula X=2piFL is invalid. NO! In the above equation, X varies when F varies. The equation does NOT mean that L varies as F varies. __________________________________________________ _______ Perhaps an example will make it clear. Suppose you have a coil which measures 1 uH at 1 MHz. It is known to have a self-resonant (parallel) frequency of 100 MHz. You measure its reactance at 1 MHz using the formula X=2*pi*F and find it to be 6.28 ohms. At 2 MHz you find it to be 12.56 ohms. At 10 MHz you find it to be 62.8 ohms. So far the reactance is changing linearly with respect to frequency. (Actually it is not perfectly linear, but the difference at these frequencies is small and probably would not be observed with run of the mill test equipment.) But, as you approach 100 MHz, you find the change is obviously no longer linear. At 95 MHz you would expect the reactance to be 6.28*95=596.6 ohms, but much to your surprise, it measures 1000 ohms. At 99 MHz, instead of the expected 6.28*99=621.72 ohms, it measures 50,000 ohms!! All the above is perfectly normal and easily observable. My point is that when a coil measures 50,000 ohms at 99 MHz, its inductance HAS TO BE L=X/(2*pi*F), or 50,000/(6.28*99)=80.4 uH! Here are your own words: "At that self-resonant frequency, the coil is behaving like a parallel resonant circuit, which of course it is, due to the parasitic capacitance between each winding." Your example ignores the capacitance, which you have stated exists. There is nothing in your formula that addresses it. You cannot use the formula or the math above (in your post) to support your point of view, because it does not contain any term for capacitance. The capacitance exists, and exhibits a larger and larger affect on the circuit as the frequency increases from 1 - 99 mHz. This is not an illusion. If you have an inductance meter which uses 99 MHz as a test frequency, it WILL MEASURE 80.4 uH. And therefore, I maintain that inductance DOES vary with frequency. How can it be otherwise? -- Bill, W6WRT |
In article ,
Paul Burridge wrote: You measure its reactance at 1 MHz using the formula X=2*pi*F and find it to be 6.28 ohms. At 2 MHz you find it to be 12.56 ohms. At 10 MHz you find it to be 62.8 ohms. So far the reactance is changing linearly with respect to frequency. (Actually it is not perfectly linear, but the difference at these frequencies is small and probably would not be observed with run of the mill test equipment.) But, as you approach 100 MHz, you find the change is obviously no longer linear. At 95 MHz you would expect the reactance to be 6.28*95=596.6 ohms, but much to your surprise, it measures 1000 ohms. At 99 MHz, instead of the expected 6.28*99=621.72 ohms, it measures 50,000 ohms!! All the above is perfectly normal and easily observable. My point is that when a coil measures 50,000 ohms at 99 MHz, its inductance HAS TO BE L=X/(2*pi*F), or 50,000/(6.28*99)=80.4 uH! This is not an illusion. If you have an inductance meter which uses 99 MHz as a test frequency, it WILL MEASURE 80.4 uH. And therefore, I maintain that inductance DOES vary with frequency. How can it be otherwise? As Spock said to Kirk, "You proceed from a false assumption." Or, to put it another way, the scenario you've just laid out contains an inherent contradiction. The inductance meter that you are using (or assuming) is not actually measuring inductance. It's measuring reactance, and back-calculating to what the inductance would be *if* it were measuring a "pure" inductance. However, as you recognize, the component that you are measuring is *not* a pure inductance. Its actual reactance is the result of interaction between its inductance, its inter-winding and distributed capacitance, and its winding resistance (at any given frequency). So, what you're observing can best be interpreted as follows: - At low frequencies (well below resonance), the component's reactance is dominated by its inductive component. It's a decent approximation of a "pure" inductance. The inductance meter gives accurate estimate of the inductive component. - At high frequencies (well above resonance), the component's reactance is dominated by its capacitive component. It becomes a decent approximation of a "pure" capacitance at some point, I suspect. At these frequencies, your simple inductance meter lies through its teeth. It "tells" you that the part's inductance is such-and- such, but it's not telling you the truth. It's hiding from you the fact that the reactance it's seeing isn't inductive at all (the reactance decreases as frequency goes up, and exhibits a capacitive phase angle). So, I think, what you're facing here is the problem which occurs when you try to force simplifying assumptions ("the component being measured is a pure inductance" and "an inductance meter actually measures inductance") outside the range in which these assumptions are valid. -- Dave Platt AE6EO Hosting the Jade Warrior home page: http://www.radagast.org/jade-warrior I do _not_ wish to receive unsolicited commercial email, and I will boycott any company which has the gall to send me such ads! |
In article ,
Paul Burridge wrote: You measure its reactance at 1 MHz using the formula X=2*pi*F and find it to be 6.28 ohms. At 2 MHz you find it to be 12.56 ohms. At 10 MHz you find it to be 62.8 ohms. So far the reactance is changing linearly with respect to frequency. (Actually it is not perfectly linear, but the difference at these frequencies is small and probably would not be observed with run of the mill test equipment.) But, as you approach 100 MHz, you find the change is obviously no longer linear. At 95 MHz you would expect the reactance to be 6.28*95=596.6 ohms, but much to your surprise, it measures 1000 ohms. At 99 MHz, instead of the expected 6.28*99=621.72 ohms, it measures 50,000 ohms!! All the above is perfectly normal and easily observable. My point is that when a coil measures 50,000 ohms at 99 MHz, its inductance HAS TO BE L=X/(2*pi*F), or 50,000/(6.28*99)=80.4 uH! This is not an illusion. If you have an inductance meter which uses 99 MHz as a test frequency, it WILL MEASURE 80.4 uH. And therefore, I maintain that inductance DOES vary with frequency. How can it be otherwise? As Spock said to Kirk, "You proceed from a false assumption." Or, to put it another way, the scenario you've just laid out contains an inherent contradiction. The inductance meter that you are using (or assuming) is not actually measuring inductance. It's measuring reactance, and back-calculating to what the inductance would be *if* it were measuring a "pure" inductance. However, as you recognize, the component that you are measuring is *not* a pure inductance. Its actual reactance is the result of interaction between its inductance, its inter-winding and distributed capacitance, and its winding resistance (at any given frequency). So, what you're observing can best be interpreted as follows: - At low frequencies (well below resonance), the component's reactance is dominated by its inductive component. It's a decent approximation of a "pure" inductance. The inductance meter gives accurate estimate of the inductive component. - At high frequencies (well above resonance), the component's reactance is dominated by its capacitive component. It becomes a decent approximation of a "pure" capacitance at some point, I suspect. At these frequencies, your simple inductance meter lies through its teeth. It "tells" you that the part's inductance is such-and- such, but it's not telling you the truth. It's hiding from you the fact that the reactance it's seeing isn't inductive at all (the reactance decreases as frequency goes up, and exhibits a capacitive phase angle). So, I think, what you're facing here is the problem which occurs when you try to force simplifying assumptions ("the component being measured is a pure inductance" and "an inductance meter actually measures inductance") outside the range in which these assumptions are valid. -- Dave Platt AE6EO Hosting the Jade Warrior home page: http://www.radagast.org/jade-warrior I do _not_ wish to receive unsolicited commercial email, and I will boycott any company which has the gall to send me such ads! |
Bill Turner wrote:
On Tue, 9 Dec 2003 19:20:25 -0500, "Ralph Mowery" wrote: The inductance is not changing. What you are measuring is not pure inductance but the coil has a stray capacitance. That is what is making the coil seof resonate. __________________________________________________ _______ I am well aware of that, but you are tap dancing around the relevance of the formula X=2*pi*F*L. Just answer this: If I have a coil of very high Q (no appreciable resistance), and I apply 100 volts of 100 MHz AC to it, and measure a current of 2 milliamps through it, then: 1. What is its reactance? 2. What is its inductance? Its impedance has been measured to have a magnitude of 50,000 ohms. If you have independent information that its Q is very high, you can assume that this impedance is made up of some combination of inductive reactance and capacitive reactance. With a single measurement such as this, that is about all you can say. It cannot be assumed to be all inductive reactance (or any particular combination of inductive and capacitive reactances), just because someone labeled the device as an inductor or because it looks like a coil. Other measurements are needed to nail the details. A parallel resonance with 50,000 ohms impedance (at some frequency) is not the same thing as an inductance with 50,000 ohms of inductive reactance (at the same frequency). They pass a similar magnitude of current at that frequency for the same applied AC, but their current phases do not match. And their reaction to nonsinusiodal waveforms is very different. -- John Popelish |
Bill Turner wrote:
On Tue, 9 Dec 2003 19:20:25 -0500, "Ralph Mowery" wrote: The inductance is not changing. What you are measuring is not pure inductance but the coil has a stray capacitance. That is what is making the coil seof resonate. __________________________________________________ _______ I am well aware of that, but you are tap dancing around the relevance of the formula X=2*pi*F*L. Just answer this: If I have a coil of very high Q (no appreciable resistance), and I apply 100 volts of 100 MHz AC to it, and measure a current of 2 milliamps through it, then: 1. What is its reactance? 2. What is its inductance? Its impedance has been measured to have a magnitude of 50,000 ohms. If you have independent information that its Q is very high, you can assume that this impedance is made up of some combination of inductive reactance and capacitive reactance. With a single measurement such as this, that is about all you can say. It cannot be assumed to be all inductive reactance (or any particular combination of inductive and capacitive reactances), just because someone labeled the device as an inductor or because it looks like a coil. Other measurements are needed to nail the details. A parallel resonance with 50,000 ohms impedance (at some frequency) is not the same thing as an inductance with 50,000 ohms of inductive reactance (at the same frequency). They pass a similar magnitude of current at that frequency for the same applied AC, but their current phases do not match. And their reaction to nonsinusiodal waveforms is very different. -- John Popelish |
In article , Bill Turner
writes: On Tue, 09 Dec 2003 04:59:29 GMT, wrote: I maintain it does. Otherwise the formula X=2piFL is invalid. NO! In the above equation, X varies when F varies. The equation does NOT mean that L varies as F varies. _________________________________________________ ________ Perhaps an example will make it clear. Suppose you have a coil which measures 1 uH at 1 MHz. It is known to have a self-resonant (parallel) frequency of 100 MHz. OK, it has a distributed capacity of 2.533 pFd. The circuit being measured is composed of a pure inductance of 1.000 uHy and pure capacitance of 2.533 pFd in parallel. We can neglect the losses in each one of those components for the sake of illustration. You measure its reactance at 1 MHz using the formula X=2*pi*F and find it to be 6.28 ohms. The inductive reactance is 6.28319 Ohms at 1 MHz and the capacitive reactance is 62.8326 KOhms at 1 MHz. In terms of susceptance, the B_sub_L is 0.159153 and 15.9153x10^-6 mhos, respectively. Total susceptance is then 0.159137 mhos or 6.28389 Ohms. [reactance meter probably isn't calibrated that close to show the slight change due to distributed capacity] At 2 MHz you find it to be 12.56 ohms. At 2 MHz, the inductive reactance is 12.5664 Ohms or 0.0795775 mhos while the distributed capacitance has 31.8306 mhos. The total susceptance is 0.0795456 mhos or 12.5714 Ohms. That is within 0.0907% of 12.56 Ohms and darn few reactance measuring gizmos are calibrated that close... At 10 MHz you find it to be 62.8 ohms. Okay, at 10 MHz, the inductive susceptance is 0.0159155 mhos and the capacitive susceptance is 159.153 x 10^-6 mhos, the total being 0.0157563 mhos or 63.4665 Ohms reactance. That's an error of 1.061% from 62.8 and still fairly reasonable for the error of a reactance meter or whatever. So far the reactance is changing linearly with respect to frequency. (Actually it is not perfectly linear, but the difference at these frequencies is small and probably would not be observed with run of the mill test equipment.) Okay, that's progress. We are agreed that test equipment can have errors...he said with a grin having worked in metrology and a standards lab for over 2 years in the past... :-) But, as you approach 100 MHz, you find the change is obviously no longer linear. Ah, but you are measuring TWO things at the same time, the parallel of the true inductor and its distributed capacity. Once you are into measuring multiple elements, you need a test setup to try to get a handle on the individual components. That is why I brought up the "true inductance" two-frequency test on a Q meter that has a calibrated tuning capacitor. That WILL establish the equivalent pure capacitor due to distributed winding capacity in the coil (the physical inductor form). Once you KNOW the distributed capacity, its just a matter of some button-pushing on a good scientific handheld calculator to derive true inductance from the reactance readings of both inductance and distributed capacity. [I recommend an HP 32 S II as an RPN fan] The parallel capacitance will definitely exist as more picoFarads in a circuit such as a FET gate which has a very high parallel resistance (or very low conductance if you can think in terms of admittance). That FET input capacitance will change the higher frequency resonance even lower. Offhand, I'd say that 2.533 pFd distributed capacity is rather high and probably is around 1.0 pFd (solenoidal type, no core)...but a FET gate input and its PCB traces to ground plane can be an additional 2.0 pFd. That's 3.0 pFd total and the self-resonance of that circuit is now 91.888 MHz. This is not an illusion. If you have an inductance meter which uses 99 MHz as a test frequency, it WILL MEASURE 80.4 uH. That "inductance meter" is still measuring TWO THINGS AT THE SAME TIME. The physical coil still has two components, the pure inductance in parallel with a pure capacitance representing the distributed capacity of the windings. Those are inseperable unless you do something like the "true inductance" test at octave separation frequencies or equivalent. A Q Meter of any kind made today, last year, or back in the pre- history before 1947, MEASURES THREE THINGS AT THE SAME TIME! Yet the Q Meter is still accurate enough to derive the equivalent parallel resistance, parallel inductance, and parallel capacitance of the physical coil's windings. [it actually measures conductance and susceptance as a total magnitude and relates that to the Q or loss factor while the calibrated frequency setting and calibrated variable capacitor allow separate "inductance" measurement even though the Q Meter is "looking" at both L and C_sub_d in parallel] ANYONE using test equipment SHOULD be aware of what their equipment does, how it works (in general), and what it really measures. Since inductance does NOT change in a passive coil (that isn't otherwise influenced by magnetic fields), what anyone measures on a particular coil is THREE THINGS: The conductance due to losses and the susceptance due to BOTH parallel inductance and parallel capacitance. Conductance will change with frequency depending on a lot of different factors (coil form, coil core, wire used, shield used (if any), dielectric of the former material, core permittivity, etc.). Susceptance will change with frequency because of the TWO components...BUT THE INDUCTIVE COMPONENT DOES NOT CHANGE. And therefore, I maintain that inductance DOES vary with frequency. How can it be otherwise? The baseline taught in all textbooks (where I learned it first) and in classes (where I learned it second) all agree that one MUST separate the components into their "pure" form and THEN derive the component parts by different tests. That is how it is perceived by most other folks based on a lot of first-principle demonstration. Inductance of a coil DOES NOT CHANGE WITH FREQUENCY. Basic definition. First-principle stuff by definition. You CAN say that APPARENT inductance changes if you are just doing one kind of test. "Apparent" isn't going to work well when this coil is dropped into a circuit thinking that "inductance changes with frequency" and the circuit contains a lot of other sneaky little components that can shoot the "apparent" reading way off. No one successfully works with L-C and active-device networks using this "apparent" reading. One separates the component parts first, then combines them into manageable parallel-equivalent or series-equivalent circuits. The ILLUSION is from looking at an impedance- or admittance-measuring instrument such as a Q Meter and thinking its calibrated inductance dial "measures inductance." It doesn't...but it comes very close. That is just the calibrated variable capacitor tuning to resonance at specific frequencies...as a convenience to the user. The capacitance markings will be accurate but any external coil that has significant parallel capacitance from its windings will add to the calibrated capacity on the dial. Some Q Meters allow variable frequency settings to do things like the octave-separation-of-frequency measurement of the external test parallel capacity. An impedance or admittance bridge type of instrument can yield different "errors" and "illusions" depending on their type/kind. Len Anderson retired (from regular hours) electronic engineer person |
All times are GMT +1. The time now is 04:44 PM. |
Powered by vBulletin® Copyright ©2000 - 2025, Jelsoft Enterprises Ltd.
RadioBanter.com