RadioBanter

RadioBanter (https://www.radiobanter.com/)
-   Homebrew (https://www.radiobanter.com/homebrew/)
-   -   S-Meter Calibration Standards? (https://www.radiobanter.com/homebrew/23767-s-meter-calibration-standards.html)

Avery Fineman November 26th 04 04:05 AM

S-Meter Calibration Standards?
 
Is there a standard RF input level per "S" Unit?

If so, please post the location. Thanks.



Ralph Mowery November 26th 04 04:25 AM


"Avery Fineman" wrote in message
...
Is there a standard RF input level per "S" Unit?

If so, please post the location. Thanks.


There is no "standard". You will often see it mentioned as 6 db of power
per S-unit. At one time 50 microvolts into the receiver was S-9 and you
went down 6 db of power per S-unit from there. I doubt that any receiver
will follow that so called "standard".




bviel November 26th 04 04:56 AM

The rule voor 50µV = S9 does still exists, this is voor HF, @50 Ohm input
impedance at receiver.
"They" (radioamateurs) use another rule for VHF.
It's in the software MultiCalc, Google will find the adres, it's for free.
Official or not, if everyone use this rule then it is a standard for me.
Greetings Bas.


"Avery Fineman" schreef in bericht
...
Is there a standard RF input level per "S" Unit?

If so, please post the location. Thanks.





James Bond November 26th 04 06:07 AM

I find that most of them seem to sort of follow the rule that 4uV is S9, 3dB
down is each S-Point, that's for VHF/UHF.

Sam

"bviel" wrote in message
...
The rule voor 50µV = S9 does still exists, this is voor HF, @50 Ohm input
impedance at receiver.
"They" (radioamateurs) use another rule for VHF.
It's in the software MultiCalc, Google will find the adres, it's for free.
Official or not, if everyone use this rule then it is a standard for me.
Greetings Bas.


"Avery Fineman" schreef in bericht
...
Is there a standard RF input level per "S" Unit?

If so, please post the location. Thanks.






---
Outgoing mail is certified Virus Free.
Checked by AVG anti-virus system (
http://www.grisoft.com).
Version: 6.0.799 / Virus Database: 543 - Release Date: 24/11/2004



Ian White, G3SEK November 26th 04 07:12 AM

Avery Fineman wrote:
Is there a standard RF input level per "S" Unit?

If so, please post the location. Thanks.

There is an IARU recommendation, which originated in Region1
(Europe/Africa) and I believe has been adopted by IARU world-wide.

The Region 1 recommendation is:
http://www.algonet.se/~k-jarl/ssa/IARU/smeter.html
Google should also bring up the full technical paper behind this, and
probably the current IARU recommendation. ISTR there's something on the
ARRL website too.

The 1990 Region 1 recommendation simply states that:

STANDARDISATION OF S-METER READINGS
1. One S-unit corresponds to a signal level difference of 6 dB,

2. On the bands below 30 MHz a meter deviation of S-9 corresponds to an
available power of -73 dBm from a continuous wave signal generator
connected to the receiver input terminals,

3. On the bands above 144 MHz this available power shall be -93 dBm,

4. The metering system shall be based on quasi-peak detection with an
attack time of 10 msec +/- 2 msec and a decay time constant of at least
500 msec.


IARU functions very much like an
Entmoot...only...not...quite...so...hoom...hasty. In 2004, they are just
starting to think about the gap between 30MHz and 144MHz:
http://home.hccnet.nl/a.dogterom/Vie...4_19_Chair.rtf


But all the other comments about S-meters are true as well: that no
S-meter actually conforms to this recommendation; that it makes no
practical difference; and that hardly anyone cares.

There's a very good web page with lots of practical measurements, at:
http://www.seed-solutions.com/gregor...rimentation/SM
eterBlues.htm


--
73 from Ian G3SEK 'In Practice' columnist for RadCom (RSGB)
http://www.ifwtech.co.uk/g3sek

Roy Lewallen November 26th 04 07:45 AM

If you're wondering if S-meters are calibrated to some standard, the
answer is that they're not. S-meters simply read the AGC voltage, which
is only approximately logarithmic. Therefore, the number of dB per S
unit typically varies from one part of the scale to another. For
example, the S-meter in my Icom 730, on 40 meters, preamp off, varies
from 1.3 to 4.0 dB per S-unit depending on where on the scale you are.
The "10 dB" increments over S-9 vary from 5.6 to 13.5 dB. Receiver
manufacturers are free to make the sensitivity whatever they want, and
seldom exceed about 5 dB per S-unit, because users complain that the
meter is too "Scotch" (insensitive) if they do.

There is a tendency for receivers to be calibrated to about 50
microvolts at one point on the meter, S-9, and there's often a
calibration adjustment for this.

Any "standard" is useless at best and misleading at worst, because it
bears no relation to what S-meters really read.

Roy Lewallen, W7EL

Avery Fineman wrote:
Is there a standard RF input level per "S" Unit?

If so, please post the location. Thanks.



J M Noeding November 26th 04 12:17 PM

On Fri, 26 Nov 2004 05:56:42 +0100, "bviel" wrote:

The rule voor 50µV = S9 does still exists, this is voor HF, @50 Ohm input
impedance at receiver.
"They" (radioamateurs) use another rule for VHF.
It's in the software MultiCalc, Google will find the adres, it's for free.
Official or not, if everyone use this rule then it is a standard for me.
Greetings Bas.


"Avery Fineman" schreef in bericht
...
Is there a standard RF input level per "S" Unit?

If so, please post the location. Thanks.




On HF there is no need for an S-meter standard because it seems to be
the rule to give report 59. If you don't you soon have a bad
reputation of giving wrong reports

I installed an MC3356 log detector and calibrated it for S9 = 50uV.
For about 15 years there has never been any signal of S9+30dB, most
reports should be around 539....579, but again you logging problems,
because everybody expect 599 and some log programs don't include the
reports, they are supposed to be 599

73
Jan-Martin
LA8AK (ex-G5BFV)
---
J. M. Noeding, LA8AK, N-4623 Kristiansand
http://home.online.no/~la8ak/c.htm

William E. Sabin November 26th 04 12:38 PM


"Avery Fineman" wrote in message
...
Is there a standard RF input level per "S" Unit?

If so, please post the location. Thanks.


My homebrew solid state receiver (see QRZ.COM) uses 5 dB per S-unit. S9
corresponds to -73 dBm available power, which is 100 microvolts open-circuit
from a 50 ohm sig gen, or 50 microvolts into a 50 ohm load. Available power
and open-circuit voltage are used because the input inpedance of the
receiver is not guaranteed to be 50 ohms. I use 5 dB per S unit because it
compresses the scale a little and it also agrees more closely with the
intuitive listening test measure that I have used for many years. The upper
end of the scale is 30 dB above S9, which is -43 dBm, a very strong signal.
Signals stronger than S9+30 dB I don't bother to measure. At S1 the signal
level is -73 - 40 = -113 dBm. Signals weaker than that S1 I don't try to
quantify.

My receiver has a low noise RF preamp with 8 dB of gain that I use on the 12
and 10 meter bands, when those bands are quiet. This makes the S meter less
accurate but I don't worry about that. To get an S meter reading I turn off
the preamp briefly.

My receiver has a custom made, computer printed scale using a calibrated sig
gen, and there are two trimpot adjustments, one for the low end and one for
the high end. This circuit uses voltage regulated opamps. The S meter
dynamics are adjusted using RC time constants.

My S meter is accurate within +/- 2 dB from 160 M to 10 M, because the
receiver is designed for this accuracy. Because of the IF and RF circuit
design, the scale calibration is fairly correct and reliable, as I
mentioned.

Bill W0IYH



William E. Sabin November 26th 04 01:05 PM


"William E. Sabin" wrote in message
...

"Avery Fineman" wrote in message
...
Is there a standard RF input level per "S" Unit?

If so, please post the location. Thanks.


My homebrew solid state receiver (see QRZ.COM) uses 5 dB per S-unit. S9
corresponds to -73 dBm available power, which is 100 microvolts
open-circuit from a 50 ohm sig gen, or 50 microvolts into a 50 ohm load.
Available power and open-circuit voltage are used because the input
inpedance of the receiver is not guaranteed to be 50 ohms. I use 5 dB per
S unit because it compresses the scale a little and it also agrees more
closely with the intuitive listening test measure that I have used for
many years. The upper end of the scale is 30 dB above S9, which is -43
dBm, a very strong signal. Signals stronger than S9+30 dB I don't bother
to measure. At S1 the signal level is -73 - 40 = -113 dBm. Signals weaker
than that S1 I don't try to quantify.

My receiver has a low noise RF preamp with 8 dB of gain that I use on the
12 and 10 meter bands, when those bands are quiet. This makes the S meter
less accurate but I don't worry about that. To get an S meter reading I
turn off the preamp briefly.

My receiver has a custom made, computer printed scale using a calibrated
sig gen, and there are two trimpot adjustments, one for the low end and
one for the high end. This circuit uses voltage regulated opamps. The S
meter dynamics are adjusted using RC time constants.

My S meter is accurate within +/- 2 dB from 160 M to 10 M, because the
receiver is designed for this accuracy. Because of the IF and RF circuit
design, the scale calibration is fairly correct and reliable, as I
mentioned.

Bill W0IYH



My receiver also has a 20 dB antenna input attenuator that can be switched
in from the front panel. This extends the upper signal range to S9 + 50 dB.
I use it very rarely.

Bill W0IYH



Ken Scharf November 26th 04 04:49 PM

On HF there is no need for an S-meter standard because it seems to be
the rule to give report 59. If you don't you soon have a bad
reputation of giving wrong reports

A signal doesn't have to be S9 to be heard 5 by 9.
I would usually tell the guy on the other end what
the S meter read, and also how readable he was.
When the band is quiet QRN wise, I could honestly
give a 59 report to someone hardly moving the meter.
OTOH with heavy QRM+QRN someone could be pumping
30db over and still be a rough copy.

Ralph Mowery November 26th 04 05:55 PM


"Ken Scharf" wrote in message
...
On HF there is no need for an S-meter standard because it seems to be
the rule to give report 59. If you don't you soon have a bad
reputation of giving wrong reports

A signal doesn't have to be S9 to be heard 5 by 9.
I would usually tell the guy on the other end what
the S meter read, and also how readable he was.
When the band is quiet QRN wise, I could honestly
give a 59 report to someone hardly moving the meter.
OTOH with heavy QRM+QRN someone could be pumping
30db over and still be a rough copy.


Then you are still passing out wrong signal reports. The first one would be
something like 5 x 2 or 5 x 3 and the second one would be 2 x 9 or 3 x 9.

The first number is how well you can understand what is being said and the
second is how strong the signal is.



Reg Edwards November 26th 04 06:36 PM

There has been a de-facto HF standard for 60 years.

The USA military first used it in specifications of radio equipment when
placing contracts with manufacturers around the end of WW2. There may have
been some restrictions on publicity at the time.

The Standard is 6 dB per S-unit and 50 micro-volts into 50 ohms at S=9.

Therefore an S-meter is essentially a power or wattmeter.

The Standard is quite logically derived.

The 6 dB fits in very nicely between a typical receiver's internal noise
level (S=0) and a typical receiver's signal overload point (S=9+30 or 40
dB).

S=9 is about half way up the scale which is linear in dB's, or S-units, from
one end to the other.

There's nothing wrong with the standard. If your S-meter reads incorrectly
then don't blame the standard - re-calibrate the meter. If you can't
re-calibrate it blame the poor quality of the meter.

I have two relatively modern commercial transceivers plus two home-brewed
transceivers. Their S-meters are accurate enough for the intended purpose.
What more should I expect?
----
Reg, G4FGQ



Avery Fineman November 26th 04 08:27 PM

In article , "William E. Sabin"
writes:

My receiver has a custom made, computer printed scale using a calibrated sig
gen, and there are two trimpot adjustments, one for the low end and one for
the high end. This circuit uses voltage regulated opamps. The S meter
dynamics are adjusted using RC time constants.

My S meter is accurate within +/- 2 dB from 160 M to 10 M, because the
receiver is designed for this accuracy. Because of the IF and RF circuit
design, the scale calibration is fairly correct and reliable, as I
mentioned.

Bill W0IYH


Thanks, Bill. I'm doing essentially the same...and expect the overall
receiver response to the flat within +/- 1 db within an octave and a
half tuning range. Accuracy of the S-Meter is only as good as the
RF level accuracy of the calibrating RF source but that's another
task and I have confidence in that. But, I have to start someplace
and that is why I asked about a "standard." I know that the U.S.
military didn't bother with any receiver S-Meter calibration standards
since around 1980, only approximate differential signal strength
readings if there was an indicator at all.



Avery Fineman November 26th 04 08:27 PM

In article , Roy Lewallen
writes:

Any "standard" is useless at best and misleading at worst, because it
bears no relation to what S-meters really read.


Roger that, and I've heard all kinds of "599" reports on-air, too. :-)

In the project I have on-going, this receiver's S-Meter (there mainly
for nostalgia purposes and because I've gotten a few old-style
microammeters that I can use) will read the average carrier power
(integrated to a time-constant yet to be settled on) at the AM
detector using a half of a quad op-amp integrator. That same
detector provides the AGC control line with appropriate DC offset
for the MC1349 gain blocks, but with (maybe) different integration
time-constants. As for the RF input to the antenna connector,
that is known within +/- 2 db down to -130 dbm by separate
calibration of my HP 608 or 606 signal generator. When
completed, this particular receiver S-Meter will be as accurate,
with a custom scale plate, as that signal generator output will
allow.

Of course, to fit the "convention" of those gratuitous "599" reports,
I could borrow from the auto industry's cheapie "gas gauge" which
has a single bulb showing "low gas" on the dashboard...using an
LED driven by an op-amp integrator-comparator for high-tech.
The indicator would show "I got signal" or "I got noise" to fit... :-)




Avery Fineman November 26th 04 08:27 PM

In article , "Ian White, G3SEK"
writes:

Is there a standard RF input level per "S" Unit?

If so, please post the location. Thanks.

There is an IARU recommendation, which originated in Region1
(Europe/Africa) and I believe has been adopted by IARU world-wide.


Thanks, Ian, and thanks to all others responding. A plus to Reg
Edwards for mentioning the U.S. military receiver specs which
I was hunting around for but could not find. :-(

Reason for asking is that I'm going to make a meter scale for a
little receiver a-building, using (nobody blanch, please) MS
Paint from a scanner (accurate 1:1) digitization of the removed
meter scale plate. I've done that with a normal-expanded scale
meter on a 120 W variable autotransformer box used on the
bench. MS Paint will do color in 256-color mode for a better
appearance. An inkjet printout on heavy photo paper stock
results in a fine-grain scale sturdy enough to replace the stock
plate in a little 2 1/2" microammeter.

Note: That works only on the old-style meters with removeable
scale plates (screw mounting type). Newer snap-together plastic
case types aren't recommended for that.

That method started on wondering how accurate an ordinary
scanner was...solved by scanning a 6-inch metal scale, printing
it, then comparing the real scale to the printout. By eyeball it
was dead-on! :-) I've done that for drill guides on PC board
stock used for both circuit boards and small enclosures since
and find it very time-economical. Machine shop accurate it isn't
but then my home shop drill locations were never that accurate
using scribe marks on "dye-chem" blue lacquer painted on
aluminum. :-) [rubber cement holds the paper printout on the
work, removes easily afterwards]



Reg Edwards November 26th 04 09:21 PM


Most people appear to be fatally attempting to use, in reverse, their
favourite S-meter as a means of calibrating the Standard. No wonder we have
so many different standards around.

The difficulties in making an accurate power meter lie solely in the very
wide range of power levels encountered.

0 to S-9 corresponds to 9 times 6 dB = 54 dB.

S-9 to +40 dB corresponds to 40 dB.

Making a very high total range of 94 dB for a power meter.

That explains why S-9 usually appears just over half way up the scale.

For a range of 94 dB it is not beyond modern technology to make a linear dB
scale out of it. The limitation is manufacturing cost. But who wants to pay
an extra hundred dollars to replace a receiver they are already happy with.

Of course, if you MUST have an accurate S-meter, the cheap way is to obtain
a blank meter scale, a fine-nibbed pen, a bottle of black ink, a signal
generator, and a 0-100 dB switched attenuator. You will proudly end up with
a work of art and a beautifully cramped scale at the bottom end. But when
done it's as accurate as you like!

Hint: There's no need to obtain a new blank scale if the existing scale can
just be turned over to its white side. Such small divergencies can make
restful breaks in between investigations of skin depths at 1 to 10 Hz of the
ocean bottom of transatlantic submarine cables.
----
Reg



Reg Edwards November 27th 04 12:26 AM

Just to tidy up.

And I've been through this before but I'm just a bloody foreigner who
favours French wine.

The S-meter is a power meter.

The standard receiver input impedance is 50 ohms.

That's why you get a conjugal match when you switch the transmit tuner from
transmit to receive but you don't get such a match when you switch back.

The standard, HF, 50 microvolts at S-9 into 50 ohms corresponds to 50
pico-watts which is an inconvenient quantity to refer to in signal strength
reports. Hence the popular S-units.

S-9 requires a standard 50-ohm signal generator, set to a standard
open-circuit 100 micro-volts, to be connected to the receiver. Receiver
manufacturers in their maintenance manuals usually prescibe this at the
non-descript frequency of 7 MHz.

The internal thermal and other noise level of a typical receiver with an
input stage consisting of a balanced modulator (the first frequency
changer), referred to the receiver's input terminals, with a receiver SSB
bandwidth of 4 KHz, is of the order of 60 dB below S-9. That is a little
less than S-zero on the meter. A signal level of the same order as the noise
takes the meter to S-zero.

A signal level of S-9 plus 40 dB, or 40 dB above 50 pico-watts, corresponds
to a signal input voltage of 50 micro-volts times 100 which equals 5
milli-volts. At which point a good receiver begins to overload and suffers
from non-linear intermod products.

Hence we have a meter range of 54 + 40 = 94 dB as displayed on a typical
meter.

All this fits in very nicely with the recognised S-meter Calibration
Standard.

(I do hope I have not made an arithmetical error. But I'm sure you Americans
get the general idea nevertheless.)
----
Reg, G4FGQ



xpyttl November 27th 04 03:30 PM

"Reg Edwards" wrote in message
...

Most people appear to be fatally attempting to use, in reverse, their
favourite S-meter as a means of calibrating the Standard. No wonder we

have
so many different standards around.


We have seen a dozen or so posters indicate that S9=50uv

Nowadays, very few rigs have analog meters. Instead, they have a computer
that can take that AGC voltage and display it however the computer decides.

In the Icom calibration procedure for the 706, there are three calibration
points known by the computer ... S0 (0 uv), S9 (50 uv), and S9+60 (50 mv).
The calibration procedure sets these three with known inputs, and presumably
the computer interpolates from there. That really isn't a bad calibration,
and I'd be surprised if other modern rigs were much different.

Now it is quite likely that there is some dependence on the AGC voltage with
frequency, but again, in a modern rig that should be manageable.

What isn't so constant, of course, is the antenna. The voltage at the
receiver could range over several orders of magnitude for the same signal
depending on the antenna. So in terms of providing input to the sender, the
S meter reading is still of limited usefulness, even if we all had
calibrated radios and agreed that S9=50uv.

...



James Bond November 27th 04 03:50 PM

The 756 pro is "properly" calibrated, each S-Point is individually
calibrated.

Sam
"xpyttl" wrote in message
...
"Reg Edwards" wrote in message
...

Most people appear to be fatally attempting to use, in reverse, their
favourite S-meter as a means of calibrating the Standard. No wonder we

have
so many different standards around.


We have seen a dozen or so posters indicate that S9=50uv

Nowadays, very few rigs have analog meters. Instead, they have a computer
that can take that AGC voltage and display it however the computer
decides.

In the Icom calibration procedure for the 706, there are three calibration
points known by the computer ... S0 (0 uv), S9 (50 uv), and S9+60 (50 mv).
The calibration procedure sets these three with known inputs, and
presumably
the computer interpolates from there. That really isn't a bad
calibration,
and I'd be surprised if other modern rigs were much different.

Now it is quite likely that there is some dependence on the AGC voltage
with
frequency, but again, in a modern rig that should be manageable.

What isn't so constant, of course, is the antenna. The voltage at the
receiver could range over several orders of magnitude for the same signal
depending on the antenna. So in terms of providing input to the sender,
the
S meter reading is still of limited usefulness, even if we all had
calibrated radios and agreed that S9=50uv.

..




---
Outgoing mail is certified Virus Free.
Checked by AVG anti-virus system (http://www.grisoft.com).
Version: 6.0.799 / Virus Database: 543 - Release Date: 25/11/2004



Ken Scharf November 27th 04 04:04 PM

Ralph Mowery wrote:
"Ken Scharf" wrote in message
...

On HF there is no need for an S-meter standard because it seems to be
the rule to give report 59. If you don't you soon have a bad
reputation of giving wrong reports


A signal doesn't have to be S9 to be heard 5 by 9.
I would usually tell the guy on the other end what
the S meter read, and also how readable he was.
When the band is quiet QRN wise, I could honestly
give a 59 report to someone hardly moving the meter.
OTOH with heavy QRM+QRN someone could be pumping
30db over and still be a rough copy.



Then you are still passing out wrong signal reports. The first one would be
something like 5 x 2 or 5 x 3 and the second one would be 2 x 9 or 3 x 9.

The first number is how well you can understand what is being said and the
second is how strong the signal is.


If you mean the actual strenth of the signal in uv at the antenna,
then you are correct. If you mean the strength of the signal in your
EARS that's another story. One is an actual measurement, the other
is subjective. (How would you measure signal strength if you
were using a receiver without an s meter, such as an old SW3?)

Ralph Mowery November 27th 04 04:30 PM


Then you are still passing out wrong signal reports. The first one

would be
something like 5 x 2 or 5 x 3 and the second one would be 2 x 9 or 3 x

9.

The first number is how well you can understand what is being said and

the
second is how strong the signal is.


If you mean the actual strenth of the signal in uv at the antenna,
then you are correct. If you mean the strength of the signal in your
EARS that's another story. One is an actual measurement, the other
is subjective. (How would you measure signal strength if you
were using a receiver without an s meter, such as an old SW3?)


I don't mean a thing. If you take some time and review the RS(T) system you
would see how it works. The first number is how well you understand what is
being said from just catching a few words to understanding everything. The
second is the strength of the signal. As the SW3 does not have an AVC
system you may be able to judge the strength by how far up you have to turn
the volume control for a certain loudness in your ears.



J M Noeding November 27th 04 05:27 PM

On Fri, 26 Nov 2004 07:05:46 -0600, "William E. Sabin"
wrote:


"William E. Sabin" wrote in message
...

"Avery Fineman" wrote in message
...
Is there a standard RF input level per "S" Unit?

If so, please post the location. Thanks.


My homebrew solid state receiver (see QRZ.COM) uses 5 dB per S-unit. S9
corresponds to -73 dBm available power, which is 100 microvolts
open-circuit from a 50 ohm sig gen, or 50 microvolts into a 50 ohm load.
Available power and open-circuit voltage are used because the input
inpedance of the receiver is not guaranteed to be 50 ohms. I use 5 dB per
S unit because it compresses the scale a little and it also agrees more
closely with the intuitive listening test measure that I have used for
many years. The upper end of the scale is 30 dB above S9, which is -43
dBm, a very strong signal. Signals stronger than S9+30 dB I don't bother
to measure. At S1 the signal level is -73 - 40 = -113 dBm. Signals weaker
than that S1 I don't try to quantify.

My receiver has a low noise RF preamp with 8 dB of gain that I use on the
12 and 10 meter bands, when those bands are quiet. This makes the S meter
less accurate but I don't worry about that. To get an S meter reading I
turn off the preamp briefly.

My receiver has a custom made, computer printed scale using a calibrated
sig gen, and there are two trimpot adjustments, one for the low end and
one for the high end. This circuit uses voltage regulated opamps. The S
meter dynamics are adjusted using RC time constants.

My S meter is accurate within +/- 2 dB from 160 M to 10 M, because the
receiver is designed for this accuracy. Because of the IF and RF circuit
design, the scale calibration is fairly correct and reliable, as I
mentioned.

Bill W0IYH



My receiver also has a 20 dB antenna input attenuator that can be switched
in from the front panel. This extends the upper signal range to S9 + 50 dB.
I use it very rarely.

Bill W0IYH


defining the proper time-constant for ssb is another problem. My
activity has mainly been VHF/UHF cw and ssb, and working a few HF
contests on cw. Found that reporting and operational style is somewhat
different on HF and VHF. While the normal report on HF is 59 or 599,
it varies much more on VHF, and 519 report is not rare on VHF -
particularly if the OP has no experience from HF.
My favourite report is 559, because it is easy to send using elbug,
and it makes some fun, particularly when everybody expects to receive
a 599 report, so the opposite OP must make a note in his log that the
report wasn't the usual type, and somebody most likely may loose score

73
Jan-Martin, LA8AK
---
J. M. Noeding, LA8AK, N-4623 Kristiansand
http://home.online.no/~la8ak/c.htm

Ralph Mowery November 27th 04 05:46 PM


defining the proper time-constant for ssb is another problem. My
activity has mainly been VHF/UHF cw and ssb, and working a few HF
contests on cw. Found that reporting and operational style is somewhat
different on HF and VHF. While the normal report on HF is 59 or 599,
it varies much more on VHF, and 519 report is not rare on VHF -
particularly if the OP has no experience from HF.
My favourite report is 559, because it is easy to send using elbug,
and it makes some fun, particularly when everybody expects to receive
a 599 report, so the opposite OP must make a note in his log that the
report wasn't the usual type, and somebody most likely may loose score

73
Jan-Martin, LA8AK


Now we are getting to my bands of operating. Instead of the RS(T) type of
reports grid squares are usually used on VHF. That eliminates the bogus 599
type of reports and while probably never used it will give a rough check to
see if you actually copied the call correct as the grids can be compaired to
the other stations that worked the same call
..



Reg Edwards November 27th 04 05:54 PM


S-meters are nothing else but power (input) meters.

Amateurs and meter manufacturers long ago learned, that when giving signal
strength reports, it is more convenient to refer to meter indications in
terms of S-units rather than micro-watts or nano-watts.

At HF, when the meter reads S-9 the power entering the receiver is 50
pico-watts. There's a slight complication above S-9 when the meter scale
changes to decibels above S-9.

When the reading is S-9 plus 40 dB the meter is actually indicating about
S-16. It's just a matter of scale graduations and printing.

The S-meter does NOT measure or even indicate field strength. It indicates
nothing except that an increase in meter reading corresponds to an increase
in field strength. Which may be nice to know but by how much of an increase
is anybody's guess.

Measured field strength depends on the type of antenna, its efficiency,
ground losses, etc. It is possible, of course, to calculate field strength
in the vicinity of the antenna from S-meter readings provided the antenna,
its directivity, transmission line, tuner and ground characteristics are all
known numerically. Which in the amateur situation they are seldom not! Or
even in the professional situation.

You've all got one. To repeat - the S-meter is a power (input) meter.
----
Reg, G4FGQ



J M Noeding November 27th 04 10:13 PM

On Sat, 27 Nov 2004 17:46:11 GMT, "Ralph Mowery"
wrote:


defining the proper time-constant for ssb is another problem. My
activity has mainly been VHF/UHF cw and ssb, and working a few HF
contests on cw. Found that reporting and operational style is somewhat
different on HF and VHF. While the normal report on HF is 59 or 599,
it varies much more on VHF, and 519 report is not rare on VHF -
particularly if the OP has no experience from HF.
My favourite report is 559, because it is easy to send using elbug,
and it makes some fun, particularly when everybody expects to receive
a 599 report, so the opposite OP must make a note in his log that the
report wasn't the usual type, and somebody most likely may loose score

73
Jan-Martin, LA8AK


Now we are getting to my bands of operating. Instead of the RS(T) type of
reports grid squares are usually used on VHF. That eliminates the bogus 599
type of reports and while probably never used it will give a rough check to
see if you actually copied the call correct as the grids can be compaired to
the other stations that worked the same call
.


the most used report on this side would be 55A in JO38XC or if you
insist on the QTH-loc it is DS80B
Somebody ask for QRA, and I believe QRA means "name of the station"


---
J. M. Noeding, LA8AK, N-4623 Kristiansand
http://home.online.no/~la8ak/c.htm

Rick Karlquist N6RK November 29th 04 05:49 PM

http://www.n6rk.com/S_unit_definitions.doc


"Avery Fineman" wrote in message
...
Is there a standard RF input level per "S" Unit?

If so, please post the location. Thanks.





Highland Ham December 1st 04 12:43 AM

Now we are getting to my bands of operating. Instead of the RS(T) type of
reports grid squares are usually used on VHF. That eliminates the bogus

599
type of reports and while probably never used it will give a rough check

to
see if you actually copied the call correct as the grids can be compaired

to
the other stations that worked the same call

=================
You are riding my hobby horse.
I find the obligatory 599 or 59 report absolute nonsense , if not stupid ,
especially in connection with contests ,because it does not provide 'any
information' . Instead one could be required (also on HF) to report the
IARU locator like for example IO87AT followed by a serial number.
However this would no doubt cause havoc among the 'mega-scorers' since it
would be much harder to copy instead of the fixed 599 or 59 , resulting
in a lower score. Although a computer database could link a callsign to
the IARU locator ,this would be more difficult, if not impossible,during a
field day or similar event.

Frank GM0CSZ / KN6WH



Highland Ham December 1st 04 12:43 AM

There has been a de-facto HF standard for 60 years.
The USA military first used it in specifications of radio equipment when
placing contracts with manufacturers around the end of WW2. There may have
been some restrictions on publicity at the time.
The Standard is 6 dB per S-unit and 50 micro-volts into 50 ohms at S=9.

=======================
I have read somewhere that it was Art Collins , of Collins Radio fame ,who
first mooted/established the above standard for up to 30 MHz. Much later
,among radio amateurs, the S9 signal level for freqs above 30 MHz was set
at 5 microvolts into 50 Ohms .

Frank GM0CSZ / KN6WH



Yuri Blanarovich December 1st 04 06:41 PM

I find the obligatory 599 or 59 report absolute nonsense , if not stupid ,
especially in connection with contests ,because it does not provide 'any
information' .


So, like you want me to sit at the rig with magnifying glass and observe the
meter with 5% accuracy?
Reports became largely redundant, most people chose to 59 them, but nobody is
stopping you from using accurate report. It just became more important in the
contest to work as many as possible, fast, rather than do scientific reporting
on signals. The other purpose is to alert the recipient that the other part of
exchange is coming.
Most established contests are using signal reports (old requirement for DXCC
QSOs). Some newer ones and VHF are not.
Check http://www.computeradio.us/TeslaCup.htm for really fair and modern HF
contest rules.

73 and 599
Yuri, K3BU.us

Noise From Afar December 1st 04 07:02 PM

The reason for the 59 or 599 is contesters don't want to waste time with
lesser reports so they always give 59 or 599
and their computers are programmed for 59 or 599
Paper loggers just enter 59 or 599 in one entry and draw a line thru the
column
Time is of the essence in a contest
This practice altho wrong has been going on for years and undoubtedly will
continue

And a contester is not after RST information per se but must satisfy the
rules -- just quick rapid contacts is the drill
Q-Rate me lads Q-Rate

--
ruido de icógnito



Someone wrote
I find the obligatory 599 or 59 report absolute nonsense , if not stupid
,
especially in connection with contests ,because it does not provide 'any
information' .




Roy Lewallen December 1st 04 10:07 PM

Yuri Blanarovich wrote:

I find the obligatory 599 or 59 report absolute nonsense , if not stupid ,
especially in connection with contests ,because it does not provide 'any
information' .



So, like you want me to sit at the rig with magnifying glass and observe the
meter with 5% accuracy?
Reports became largely redundant, most people chose to 59 them, but nobody is
stopping you from using accurate report. . .


It's lots of fun to jump into a DX contest for a while and give honest
RST reports. Totally blows the other guy's mind. I suspect most can't
even figure out a way to log it. I'm sure Yuri would get a big kick out
of folks doing that to him during a hot contest.

Roy Lewallen, W7EL

Allodoxaphobia December 1st 04 10:13 PM

On Wed, 1 Dec 2004 11:02:27 -0800, Noise From Afar hath writ:
The reason for the 59 or 599 is contesters don't want to waste time with
lesser reports so they always give 59 or 599
and their computers are programmed for 59 or 599
Paper loggers just enter 59 or 599 in one entry and draw a line thru the
column
Time is of the essence in a contest
This practice altho wrong has been going on for years and undoubtedly will
continue

And a contester is not after RST information per se but must satisfy the
rules -- just quick rapid contacts is the drill
Q-Rate me lads Q-Rate


That's whay (as an earlier poster in the thread mentioned) VHF contests
are more honest. The exchange has to be the (correct) grid square.

Thus, the contest judges have a excellent way to locate mis-matched
calls:grids.

73
Jonesy
--
| Marvin L Jones | jonz | W3DHJ | linux
| Gunnison, Colorado | @ | Jonesy | OS/2 __
| 7,703' -- 2,345m | config.com | DM68mn SK

J M Noeding December 1st 04 11:17 PM

On Wed, 1 Dec 2004 11:02:27 -0800, "Noise From Afar"
wrote:

The reason for the 59 or 599 is contesters don't want to waste time with
lesser reports so they always give 59 or 599
and their computers are programmed for 59 or 599
Paper loggers just enter 59 or 599 in one entry and draw a line thru the
column


that is bad logging practice,
you write 5 for 559, 7 for 579, and leave open for 599

with 25WPM it is no real problem on this side
---
J. M. Noeding, LA8AK, N-4623 Kristiansand
http://home.online.no/~la8ak/c.htm

Ralph Mowery December 1st 04 11:36 PM


"Noise From Afar" wrote in message
news:7zord.190863$hj.186448@fed1read07...
The reason for the 59 or 599 is contesters don't want to waste time with
lesser reports so they always give 59 or 599
and their computers are programmed for 59 or 599
Paper loggers just enter 59 or 599 in one entry and draw a line thru the
column
Time is of the essence in a contest
This practice altho wrong has been going on for years and undoubtedly will
continue

And a contester is not after RST information per se but must satisfy the
rules -- just quick rapid contacts is the drill
Q-Rate me lads Q-Rate


AS time and the Q rate is important and everyone is going 599, why don't
the contest rules just do away with that requirement ? What good is it ?



Yuri Blanarovich December 2nd 04 03:24 AM


It's lots of fun to jump into a DX contest for a while and give honest
RST reports. Totally blows the other guy's mind. I suspect most can't
even figure out a way to log it. I'm sure Yuri would get a big kick out
of folks doing that to him during a hot contest.

Roy Lewallen, W7EL



No problem, just one more keystroke, unless one has jumped the gun, assumed 599
and logged the QSO before the other station sent the real report. Then it is
few more keystrokes :-(

Yuri, K3BU.us


All times are GMT +1. The time now is 03:47 PM.

Powered by vBulletin® Copyright ©2000 - 2025, Jelsoft Enterprises Ltd.
RadioBanter.com