Richard,
What you said is largely accurate, however at low S/N ratios, or where the distortion becomes comparable to the signal level, the reading of the composite signal (signal+noise+distortion) with anything other than an RMS meter could produce erroneous results. Joe W3JDR "Richard Clark" wrote in message ... On Fri, 14 Oct 2005 13:08:22 -0500, "Steve Nosko" wrote: A meter, a pure 1kHz tone modulated signal generator and a 1kHz notch is all that is needed. What happens if you don't have a "real" RMS meter? I don't know. Hi Steve, You don't need a "real" RMS meter. The expressed requirement for a pure 1kHz tone provides the necessary sine wave shape such that it simply becomes a matter of scale calibration. If you had said a square wave 1KHz tone (nothing pure about that), then you would have to dig deep for a "real" RMS meter. That too, could be scaled, but I wouldn't count on it because it would be a rare amplifier chain that could faithfully keep it square - and the notch would inject it into the measurement as distortion and noise. 73's Richard Clark, KB7QHC |