When you switch in an extra 10 db of attenuation at the input, does
the spur drop 10 db as well as the wanted signal drop 10 db? If spur drops more then IM is being generated in the analyzer from overload possibly. 73 Gary K4FMX On Tue, 23 Nov 2004 12:38:05 -0000, "zindazenda" wrote: hello... my old but trusty hp141t (1.2ghz) appears to have developed a fault/out of calibration?? it all works but now there is a spurious reading (peak) about 6.5mhz down from the fundamental peak..about 55db down..its always there when a signal is present (but not of course when no input.) It's defiantly the analyser as I have tried different tx sources and its always present.. sniffing rf via bird 43 attenuator.. Any ideas?? simon g0zen |