The instrument you have goes with a GR1710 vector network analyzer
system, rated 400kHz-500MHz.
It is likely 50 ohms as the 75 ohm versions were clearly marked "75
ohms" all over the place.
RF IN is from the VNA sweeper,
REFLECT OUT is the reflected signal from the device under test,
REFERENCE OUT is the reference signal from the VNA
UNKNOWN is the DUT port.
Just a directional bridge arrangement, but set up for coherent
(vector, phase-referenced) measurements.