![]() "Richard Clark" wrote in message ... On Fri, 14 Oct 2005 13:08:22 -0500, "Steve Nosko" wrote: A meter, a pure 1kHz tone modulated signal generator and a 1kHz notch is all that is needed. What happens if you don't have a "real" RMS meter? I don't know. Hi Steve, You don't need a "real" RMS meter. The expressed requirement for a pure 1kHz tone provides the necessary sine wave shape such that it simply becomes a matter of scale calibration. If you had said a square wave 1KHz tone (nothing pure about that), then you would have to dig deep for a "real" RMS meter. That too, could be scaled, but I wouldn't count on it because it would be a rare amplifier chain that could faithfully keep it square - and the notch would inject it into the measurement as distortion and noise. 73's Hi Richard, I don'r know about that. For the un-notched signal, yes, where the dominant component is the sine wave. However, not knowing how a non-RMS meter may respond to the notched-out (predomanantly noise) signal, I'd thing there is a possible cause for error compared to an RMS meter. 73, Steve, K,9.D;C'I |